Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: Neuromuscular monitoring by acceleromyography assesses the effects of non-depolarising neuromuscular blocking agents used during anaesthesia induction to optimise intubation conditions. A new type of neuromuscular monitor, TOF-Cuff, integrates electrode stimulation into a blood pressure monitoring cuff. Comparisons of this device with TOF-Scan, considered a clinical standard acceleromyography device, have not been published.
Material And Methods: This prospective, observational study was approved by the Ethics Committee East Switzerland (BASEC-nr. 2016-02044), and patients' consent was obtained before inclusion. The study's aim was to compare TOF-Cuff with TOF-Scan by measuring the duration from the administration of a neuromuscular blocking agent to a train-of-four (TOF) ratio of 0%. After anaesthesia induction, atracurium was administered (0.5 mg kg-1) and TOF ratios were recorded every 15 seconds using the two devices simultaneously. Patients were grouped according to body mass index (< or ≥ 30 kg m-2).
Results: Twenty-five non-obese and twenty-five obese patients were included. In non-obese patients, bias was -3 s (± 21.2; limits of agreement -44.7 to 38.4; P = 0.702). In obese patients, bias was -20 s (± 35.0; limits of agreement -88.6 to 48.6; P = 0.0139). Large intra-individual differences of up to 60 seconds were detected even in non-obese patients.
Conclusions: A significant systematic difference in the time to reach a TOF ratio of 0% was found when using the two devices in obese patients. In non-obese and obese patients, there were large intra-individual and clinically relevant differences. The two devices cannot be used interchangeably.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10172956 | PMC |
http://dx.doi.org/10.5114/ait.2020.98124 | DOI Listing |