Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Study Design: Retrospective cohort study.
Objective: We intend to evaluate the accuracy and safety of cervical pedicle screw (CPS) insertion under O-arm-based 3-dimensional (3D) navigation guidance.
Methods: This is a retrospective study of patients who underwent CPS insertion under intraoperative O-arm-based 3D navigation during the years 2009 to 2018. The radiological accuracy of CPS placement was evaluated using their intraoperative scans.
Results: A total of 297 CPSs were inserted under navigation. According to Gertzbein classification, 229 screws (77.1%) were placed without any pedicle breach (grade 0). Of the screws that did breach the pedicle, 51 screws (17.2%) had a minor breach of less than 2 mm (grade 1), 13 screws (4.4%) had a breach of between 2 and 4 mm (grade 2), and 4 screws (1.3%) had a complete breach of 4 mm or more (grade 3). Six screws were revised intraoperatively. There was no incidence of neurovascular injury in this series of patients. 59 of the 68 breaches (86.8%) were found to perforate laterally, and the remaining 9 (13.2%) medially. It was noted that the C5 cervical level had the highest breach rate of 33.3%.
Conclusions: O-arm-based 3D navigation can improve the accuracy and safety of CPS insertion. The overall breach rate in this study was 22.9%. Despite these breaches, there was no incidence of neurovascular injury or need for revision surgery for screw malposition.