The ASTA MicroIDSys system (ASTA, Suwon, Korea) is a newly developed Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) system for identification of microorganisms. We compared the performance of the ASTA MicroIDSys system with that of the VITEK MS system (bioMérieux, Marcy l'Etoile, France) for identifying clinical microorganisms. A total 2055 isolates including 1910 bacteria and 145 yeasts were tested. Among them, the VITEK MS correctly identified 1999 (97.3%) isolates to species level and 26 (1.3%) to the genus level. The ASTA MicroIDSys correctly identified 1988 (96.7%) isolates to species level and 28 (1.4%) to the genus level. The VITEK MS and ASTA MicroIDSys misidentified one isolate and four (0.2%) isolates, respectively, and provided no identification for 29 (1.4%) and 35 (1.7%) isolates, respectively. The performance of the ASTA MicroIDSys was comparable to that of the VITEK MS for identification of clinically relevant bacterial and yeast isolates.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jiac.2020.08.004DOI Listing

Publication Analysis

Top Keywords

asta microidsys
24
matrix-assisted laser
8
laser desorption/ionization
8
desorption/ionization time-of-flight
8
time-of-flight mass
8
mass spectrometry
8
microidsys system
8
performance asta
8
correctly identified
8
isolates species
8

Similar Publications

Clinical Evaluation of VITEK MS PRIME with PICKME Pen for Bacteria and Yeasts, and RUO Database for Filamentous Fungi.

Microorganisms

May 2024

Department of Laboratory Medicine, International St. Mary's Hospital, College of Medicine, Catholic Kwandong University, Incheon 22711, Republic of Korea.

The VITEK MS PRIME (bioMérieux, Marcy-l'Étoile, France), a newly developed matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system, alongside the VITEK PICKME pen (PICKME), offers easy sample preparation for bacteria and yeasts. The VITEK MS PRIME also offers two software platforms for filamentous fungi: the IVD database and the RUO database. Our study evaluated its identification agreement on 320 clinical isolates of bacteria and yeasts, comparing PICKME and traditional wooden toothpick sampling techniques against MicroIDSys Elite (ASTA) results.

View Article and Find Full Text PDF

Purpose: species can cause infections, and associations with cancer are being increasingly reported. As their clinical significance differs, accurate identification of individual species is important. However, matrix-assisted laser desorption/ionization-time of flight mass spectrometry has not been found to be effective in identifying species in previous studies.

View Article and Find Full Text PDF

Identification of the rare yeast Cutaneotrichosporon (Trichosporon) debeurmannianum from diabetic foot infection.

J Clin Lab Anal

December 2022

Department of Laboratory Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.

Background: Diabetic foot infection is the most common complications of diabetes mellitus. Although most of the diabetic foot infections has been known to be caused by aerobic and anaerobic bacteria, mycotic diabetic foot infection caused by Candida species has also been reported recently. Here, we present the first case of diabetic foot infection caused by Cutaneotrichosporon debeurmannianum (previously known as Trichosporon debeurmannianum).

View Article and Find Full Text PDF

Filamentous fungi are a major cause of life-threatening infections in immunocompromised patients; thus, rapid and accurate identification is critical. Filamentous fungal identification by matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) has been demonstrated with high sensitivity and reproducibility; however, its wider application has been limited in clinical laboratories because of practical challenges such as database availability or lack of standardization. In this study, we compared the performance of the Bruker Biotyper, ASTA MicroIDSys, and Vitek MS for 84 clinical filamentous fungal isolates.

View Article and Find Full Text PDF

This study compared the accuracy of a new MALDI-TOF mass spectrometry system, ASTA MicroIDSys system, with that of MALDI Biotyper system for the identification of reference and clinical bacterial and yeast strains. The identification accuracy of the 2 systems was compared using a total of 406 strains comprising 142 aerobic and 180 anaerobic bacterial strains and 84 yeast strains. The genus and species identification rates were 98.