Each year, 3.4 million people die from waterborne diseases worldwide. Development of a rapid and portable platform for detecting and monitoring waterborne pathogens would significantly aid in reducing the incidence and spread of infectious diseases. By combining optical methods and smartphone technology with molecular assays, the sensitivity required to detect exceedingly low concentrations of waterborne pathogens can readily be achieved. Here, we implement smartphone-based particle diffusometry (PD) detection of loop-mediated isothermal amplification (LAMP) targeting the waterborne pathogen Vibrio cholerae (V. cholerae). By measuring the diffusion of 400 nm streptavidin-coated fluorescent nanoparticles imaged at 68X magnification on a smartphone, we can detect as few as 6 V. cholerae cells per reaction (0.66 aM) in just 35 minutes. In a double-blinded study with 132 pond water samples, we establish a 91.8% sensitivity, 95.2% specificity, and 94.3% accuracy of the smartphone-based PD platform for detection of V. cholerae. Together, these results demonstrate the utility of this smartphone-based PD platform for rapid and sensitive detection of V. cholerae at the point of use.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7532658 | PMC |
http://dx.doi.org/10.1016/j.bios.2020.112497 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!