Purpose: In this study, we aimed at investigating the impact of deterministic signal loss on image quality and, thus, on optical coherence tomography angiography (OCTA) measurements performed by the RTVue-XR Avanti System.

Methods: Absorptive filters with different optical densities (ODs) were used to simulate signal loss in a controlled way in 30 eyes from 15 healthy subjects. Scan Quality (SQ), provided by the AngioVue software, was applied as a measure of image quality.

Results: Assessing the effect of decreased light transmittance on SQ values as well as that of attenuated image quality on OCTA parameters revealed a strong systematic dependence between these quantities. Attenuated image quality was associated with significantly decreased macular and peripapillary vessel density (VD) values, and we calculated a correction factor of 2.27% to 3.97% for a one-unit change in SQ for the different VD parameters.

Conclusions: Our findings suggest that the influence of systematic changes in image quality on OCTA parameters needs to be considered during patient follow-up in order to make valid assessment of progression.

Translational Relevance: For accurate evaluation of longitudinal changes in OCTA parameters, equal scan quality or using a correction factor is suggested.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7401964PMC
http://dx.doi.org/10.1167/tvst.9.5.10DOI Listing

Publication Analysis

Top Keywords

image quality
16
signal loss
12
octa parameters
12
impact deterministic
8
deterministic signal
8
scan quality
8
attenuated image
8
quality octa
8
correction factor
8
quality
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!