Background: Due to the low concentration of androgens in women and the limitation of immunoassays, it remains a challenge to accurately determine the levels of serum androgens in polycystic ovary syndrome (PCOS) patients for clinical laboratories. In this report, a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated for simultaneous quantitation of testosterone (T), androstenedione (A4), dehydroepiandrosterone sulfate (DHEAS), dihydrotestosterone (DHT), and 17-hydroxyprogesterone (17-OHP) that are associated with PCOS.
Methods: The serum samples were processed by protein precipitation and solid phase extraction before analysis with the in-house developed LC-MS/MS. The chromatographic separation was achieved with a C18 column, using a linear gradient elution with two mobile phases: 0.02% formic acid in water (phase A) and 0.1% formic acid in methanol (phase B). The separated analytes were detected by positive or negative electrospray ionization mode under multiple reaction monitoring (MRM).
Results: The assay for all the five analytes was linear, stable, with imprecision less than 9% and recoveries within ±10%. The lower limits of quantification were 0.05, 0.05, 5, 0.025, and 0.025 ng/mL for T, A4, DHEAS, DHT, and 17-OHP, respectively. In the receiver operating characteristic curve (ROC) analyses with the PCOS (n = 63) and healthy (n = 161) subjects, the AUC of the four-androgen combined was greater than that of any single androgen tested in PCOS diagnosis.
Conclusions: The LC-MS/MS method for the four androgens and 17-OHP showed good performance for clinical implementation. More importantly, simultaneous quantitation of the four androgens provided better diagnostic power for PCOS.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7755789 | PMC |
http://dx.doi.org/10.1002/jcla.23539 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!