Background: The O-arm O2 imaging system (OAO2) is an intraoperative cone beam 3D tomogram imaging tool with a wide enough field of view to perform intraoperative fiducial registration with standard stereotactic frames. However, the OAO2 3D images (cone beam CT) provide limited tissue contrast, which may reduce the accuracy of fusion to a preoperative targeting MRI for planning awake deep brain stimulation (DBS) surgeries. Therefore, most users obtain a preoperative CT scan to use as the reference exam for computational fusion with the preoperative targeting MRI and the intraoperative OAO2 cone beam CT.
Objective: In this study, we retrospectively analyzed the discrepancy between stereotactic coordinates of deep brain targets on MRI derived from intraoperative OAO2 fiducial registration with and without the use of preoperative CT as the reference for image fusion.
Methods: Preoperative stereotactic CT/MRI and intraoperative OAO2 cone beam CT were retrospectively evaluated for 27 consecutive DBS patients, using two commercial surgical planning software packages (BrainLab Elements and Medtronic Stealth 8). The anterior commissure, posterior commissure, and left subthalamic nucleus were identified on preoperative MRI. Each patient had intraoperative fiducial registration using the OAO2 with a Leksell headframe. For each subject, the reference scan for image fusion was set as either the preoperative CT or the preoperative MRI (volumetric T1 with contrast). Computed stereotactic coordinates for each target were then compared.
Results: For 8 of 27 subjects, a discrepancy greater than 1.0 mm for at least one designated target was observed utilizing the Medtronic Stealth S8 planning station when a preoperative CT scan was not used. An additional 5 (5/27) had a discrepancy greater than 2 mm. The most common discrepancy was in the z axis. No coordinate discrepancies greater than 1 mm were observed utilizing BrainLab Elements.
Conclusions: Caution is advised in fusing intraoperative OAO2 images directly to preoperative MRI without a preoperative CT as the reference exam for image fusion, as the specific fusion algorithm employed may unpredictably affect targeting accuracy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1159/000509312 | DOI Listing |
J Contemp Dent Pract
September 2024
RAK College of Dental Sciences, RAK Medical and Health Sciences University, Ras Al Khaimah, United Arab Emirates.
Aim: The aim of this cross-sectional prospective study was to evaluate the bone density changes around the bicortical corticobasal implant placed in the maxilla over 18 months of follow-up using cone-beam computed tomography (CBCT), focusing on the comparison between the anterior and posterior teeth and regions.
Materials And Methods: Thirty-five subjects (20, 53.26%, were males, and 15, 46.
J Contemp Dent Pract
September 2024
Department of Pediatrics Dentistry and Orthodontics, Faculty Odonto-Stomatology, Can Tho University of Medicine and Pharmacy, Can Tho City, Vietnam.
Aim: This study aimed to evaluate the impact of a combination of immediate implant placement with maxillary sinus augmentation (MSA) solely using platelet-rich fibrin (PRF) on guided bone regeneration.
Materials And Methods: An interventional before-after (pre-post) study design was used with 30 dental patients (≥18 years of age; 14 males and 16 females) with initial bone heights ranging between 4 and 6 mm. Following the general check-up and the creation of a study model, the planned implant location demonstrated an external right maxilla diameter of more than 5 mm, thereby validating the cone-beam computed tomography (CBCT) radiograph.
J Oral Rehabil
January 2025
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Ankara Medipol University, Ankara, Turkey.
Background: The foramen of Huschke (FH) and the petrotympanic fissure (PTF) are anatomical structures that can influence temporomandibular joint disorders (TMD) by potentially affecting the movement and function of the mandibular condyle.
Objective: This study investigates the relationship between patients with TMD and the presence of FH and PTF to enhance diagnostic and therapeutic approaches.
Methods: This retrospective study analyzed cone beam computed tomography (CBCT) images from 212 patients.
Sci Rep
January 2025
Office of Research Affairs, Faculty of Dentistry, Chulalongkorn University, Bangkok, 10330, Thailand.
Although the modulus of elasticity of the human periodontal ligament (E) values used in dentistry widely ranged from 0.01 to 175 MPa, the exact E value has not been determined. This study aimed to verify whether and how E values affect the stress distribution over the tooth and periodontium structures, and to determine the appropriate E range.
View Article and Find Full Text PDFInt J Oral Maxillofac Surg
January 2025
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand. Electronic address:
The aim of this study was to determine the effects of mandibular setback surgery exceeding 5 mm on upper airway and sleep quality in skeletal Class III patients, with comparisons to Class I controls. Sixteen individuals per group were selected based on their ANB angle and surgical need. 2D and 3D airway analyses were conducted.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!