Introduction: The aim of this study was to evaluate the cyclic fatigue resistance of different heat-treated nickel-titanium reciprocating instruments in 2 different situations: new and used instruments after preparing 3 curved canals.

Methods: A total sample of 60 nickel-titanium instruments of 3 systems (n = 20 per system) were used in this study: ProDesign R (tip 25, 0.06 taper; Easy Dental Equipment, Belo Horizonte, Brazil), Reciproc Blue (tip 25, 0.08v taper; VDW, Munich, Germany), and WaveOne Gold (tip 25, 0.07v taper; Dentsply Sirona, Ballaigues, Switzerland). Thirty new instruments (n = 10 per system) were used to prepare 90 curved single-rooted mandibular premolars (n = 30). Each instrument was used to prepare 3 root canals, and after each canal preparation the instrument was ultrasonic cleaned and submitted to autoclave sterilization procedures. The other 30 instruments (n = 10 per system) were kept without use. Then, the new and used instruments were subjected to the cyclic fatigue test in an artificial canal with a 30° angle and a 5-mm radius of curvature. The time and number of cycles to fatigue were recorded. Data were analyzed using the unpaired t test for intragroup comparison. For intergroup comparison, analysis of variance and the Tukey test for multiple comparisons were used.

Results: The intergroup comparison of new instruments showed that ProDesign R had the highest cyclic fatigue resistance followed by Reciproc Blue and WaveOne Gold (P < .05). Regarding the used instruments, WaveOne Gold had the lowest cyclic fatigue resistance (P < .05). The intragroup comparison between new and used instruments showed that WaveOne Gold and ProDesign R presented a significant reduction in the cyclic fatigue resistance after simulated clinical use (P < .05); no difference was found with the Reciproc Blue instruments (P > .05).

Conclusions: ProDesign R had the highest cyclic fatigue resistance, whereas WaveOne Gold had the lowest for new and used instruments. Simulated clinical use affected the cyclic fatigue resistance of ProDesign R and WaveOne Gold but not of Reciproc Blue instruments.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.joen.2020.08.010DOI Listing

Publication Analysis

Top Keywords

cyclic fatigue
32
fatigue resistance
28
waveone gold
24
reciproc blue
16
instruments
13
simulated clinical
12
cyclic
8
nickel-titanium reciprocating
8
reciprocating instruments
8
instruments simulated
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!