Background: The cementation of a new liner into a well-fixed acetabular component is common during revision total hip arthroplasty (THA) for many indications, but most commonly for lack of a modern, compatible, highly cross-linked polyethylene (HXLPE) liner. However, little is known about the intermediate-term to long-term durability of this strategy. The purpose of this study was to evaluate the implant survivorship, risk of complications, clinical outcomes, and radiographic results of cementing a new HXLPE liner into a well-fixed acetabular component.
Methods: We retrospectively identified 323 revision THAs in which a nonconstrained HXLPE liner was cemented into a well-fixed acetabular component. The mean age at the time of the revision THA was 63 years, and 50% of patients were female. The most common indications for revision THA were polyethylene wear and osteolysis (48%), aseptic femoral loosening (35%), and hip instability (8%). The mean follow-up was 9 years.
Results: Polyethylene liner failure occurred in 11 cases (3%). In all cases, the cemented liner dissociated from the acetabular component. At 10 years, the survivorship free from any revision was 80% (95% confidence interval [CI], 75% to 84%) and the survivorship free from any reoperation was 77% (95% CI, 72% to 82%). The most common reason for re-revision was dislocation (45% of reoperations). A dislocation occurred in 17% of cases. Hips that underwent revision for instability were significantly more likely to dislocate compared with hips that underwent revision for liner wear (hazard ratio [HR], 2.3 [95% CI, 1.2 to 4.5]; p = 0.02). Elevated rim or face-changing liners were significantly more likely to dissociate than flat liners (HR, 9.0 [95% CI, 1.2 to 70.6]; p = 0.04).
Conclusions: Cementation of a nonconstrained HXLPE liner into a well-fixed acetabular component during revision THA provided durable fixation with only a small number of failures at the cement interface (3%). Instability after this procedure remains a concern, but this is multifactorial in nature. These data support the continued use of this technique, when necessary, during revision THA.
Level Of Evidence: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.2106/JBJS.19.01441 | DOI Listing |
Cureus
October 2024
Trauma and Orthopedics, Royal Orthopaedic Hospital, Birmingham, GBR.
Pseudotumors are a rare complication of total hip arthroplasty (THA), arising from local soft tissue reactions. These reactions can lead to painful joint effusions and prosthetic loosening, often necessitating revision surgery. Metal-on-metal and metal-on-polyethylene prostheses are particularly prone to this complication due to the accumulation of metal debris from prosthetic wear, which represents a significant drawback.
View Article and Find Full Text PDFHSS J
November 2024
The Stavros Niarchos Foundation Complex Joint Reconstruction Center, Hospital for Special Surgery, New York, NY, USA.
Polyethylene liner exchange and bone grafting is an effective surgical option for the management of periacetabular osteolysis following total hip arthroplasty with well-fixed cups and intact liner locking mechanisms. We aimed to evaluate the revision-free survivorship and radiographic lesion progression after polyethylene liner exchange and bone grafting is performed for periacetabular osteolysis. A systematic review of the literature was performed.
View Article and Find Full Text PDFAnn Jt
May 2024
Department of Surgery, The University of Auckland, Auckland, New Zealand.
Removing well-fixed uncemented components can be challenging. With thoughtful surgical planning, appropriate surgical instruments, and proper surgical techniques, most implants can be removed expeditiously with little bone loss and minimal impact on the subsequent reconstruction. Preoperative planning is one of the most essential steps to remove uncemented implants.
View Article and Find Full Text PDFBone Jt Open
June 2024
Exeter Hip Group, Exeter Hip Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon University Healthcare NHS Foundation Trust, Exeter, UK.
Aims: In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.
View Article and Find Full Text PDFAims: There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!