Background: Diffusion-weighted imaging (DWI) is common for evaluating pediatric musculoskeletal lesions, but suffers from geometric distortion and intense acoustic noise.

Purpose: To investigate the performance of a near-silent and distortion-free DWI sequence (DW-SD) relative to standard echo-planar DWI (DW-EPI) in pediatric extremity MRI.

Study Type: Prospective validation study.

Subjects: Thirty-nine children referred for extremity MRI.

Field Strength/sequence: DW-EPI and DW-SD, based on a rotating ultrafast sequence modified with sinusoidal diffusion preparation gradients, at 3T.

Assessment: DW-SD image quality (S ) was assessed from 0 (nondiagnostic) to 5 (outstanding) and comparative image quality (S ) (from -2 = DW-EPI more delineated to +2 = DW-SD more delineated, 0 = same). ADC measured by DW-SD and DW-EPI were compared in bone marrow, muscle, and lesions.

Statistical Tests: Wilcoxon rank-sum test and confidence interval of proportions (CIOP) were calculated for S , Student's t-test, coefficient of variation (COV), and Bland-Altman analysis for ADC values, and intraclass correlation coefficient (ICC) for interreader agreement.

Results: DW-SD and DW-EPI ADC values for bone marrow, muscle, and lesions were not significantly different (P = 0.3, P = 0.2, and P = 0.27, respectively) and had an overall ADC COV of 14.8% (95% confidence interval: 12.3%, 16.9%) and no significant proportional bias on Bland-Altman analysis. S CIOP was rated diagnostic or better (score of 3, 4, or 5) in 72-98% of cases for bone marrow, muscle, and soft tissues. DW-SD was equivalent to or preferred over DW-EPI in muscles and soft tissues, with CIOP 86-93% and 93%, respectively. Lesions were equally visualized on DW-SD and DW-EPI in 40-51%, with DW-SD preferred in 44-56% of cases. DW-SD was rated significantly better than DW-EPI across all comparative variables that included bone marrow, muscle, soft tissue, cartilage, and lesions (P < 0.05). Readers had moderate to near-perfect (ICC range = 0.45-0.85).

Data Conclusion: DW-SD of the extremities provided similar ADC values and improved image quality compared with conventional DW-EPI. Level of Evidence 2 Technical Efficacy Stage 2 J. MAGN. RESON. IMAGING 2021;53:504-513.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7856171PMC
http://dx.doi.org/10.1002/jmri.27330DOI Listing

Publication Analysis

Top Keywords

bone marrow
16
marrow muscle
16
dw-sd dw-epi
12
dw-sd
10
near-silent distortion-free
8
pediatric musculoskeletal
8
dw-epi
8
image quality
8
confidence interval
8
bland-altman analysis
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!