Purpose: Cardiac stimulation (CS) limits to gradient coil switching speed are difficult to measure in humans; instead, current regulatory guidelines (IEC 60601-2-33) are based on animal experiments and electric field-to-dB/dt conversion factors computed for a simple, homogeneous body model. We propose improvement to this methodology by using more detailed CS modeling based on realistic body models and electrophysiological models of excitable cardiac fibers.

Methods: We compute electric fields induced by a solenoid, coplanar loops, and a commercial gradient coil in two human body models and a canine model. The canine simulations mimic previously published experiments. We generate realistic fiber topologies for the cardiac Purkinje and ventricular muscle fiber networks using rule-based algorithms, and evaluate CS thresholds using validated electrodynamic models of these fibers.

Results: We were able to reproduce the average measured canine CS thresholds within 5%. In all simulations, the Purkinje fibers were stimulated before the ventricular fibers, and therefore set the effective CS threshold. For the investigated gradient coil, simulated CS thresholds for the x-, y-, and z-axis were at least one order of magnitude greater than the International Electrotechnical Commission limit.

Conclusion: We demonstrate an approach to simulate gradient-induced CS using a combination of electromagnetic and electrophysiological modeling. Pending additional validation, these simulations could guide the assessment of CS limits to MRI gradient coil switching speed. Such an approach may lead to less conservative, but still safe, operation limits, enabling the use of the maximum gradient amplitude versus slew rate parameter space of recent, powerful gradient systems.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7722025PMC
http://dx.doi.org/10.1002/mrm.28472DOI Listing

Publication Analysis

Top Keywords

gradient coil
16
body models
12
cardiac stimulation
8
stimulation limits
8
limits mri
8
mri gradient
8
electromagnetic electrophysiological
8
coil switching
8
switching speed
8
gradient
7

Similar Publications

Purpose: To propose a domain-conditioned and temporal-guided diffusion modeling method, termed dynamic Diffusion Modeling (dDiMo), for accelerated dynamic MRI reconstruction, enabling diffusion process to characterize spatiotemporal information for time-resolved multi-coil Cartesian and non-Cartesian data.

Methods: The dDiMo framework integrates temporal information from time-resolved dimensions, allowing for the concurrent capture of intra-frame spatial features and inter-frame temporal dynamics in diffusion modeling. It employs additional spatiotemporal ($x$-$t$) and self-consistent frequency-temporal ($k$-$t$) priors to guide the diffusion process.

View Article and Find Full Text PDF

Fast Hadamard-Encoded 7T Spectroscopic Imaging of Human Brain.

Tomography

January 2025

NextGen Precision Health, Department of Radiology, University of Missouri Columbia, 1030 Hitt Street, Columbia, MO 65201, USA.

: The increased SNR available at 7T combined with fast readout trajectories enables accelerated spectroscopic imaging acquisitions for clinical applications. In this report, we evaluate the performance of a Hadamard slice encoding strategy with a 2D rosette trajectory for multi-slice fast spectroscopic imaging at 7T. : Moderate-TE (~40 ms) spin echo and J-refocused polarization transfer sequences were acquired with simultaneous Hadamard multi-slice excitations and rosette in-plane encoding.

View Article and Find Full Text PDF

Background: Quantitative Susceptibility Mapping (QSM) provides a non-invasive post-processing method to investigate alterations in magnetic susceptibility (χ), reflecting iron content within brain regions implicated in neurodegenerative diseases (NDDs).

Purpose: To investigate alterations in thalamic χ in patients with NDDs using QSM.

Study Type: Systematic review and meta-analysis.

View Article and Find Full Text PDF

Magnetic nanoparticle hyperthermia (MNH) emerges as a promising therapeutic strategy for cancer treatment, leveraging alternating magnetic fields (AMFs) to induce localized heating through magnetic nanoparticles. However, the interaction of AMFs with biological tissues leads to non-specific heating caused by eddy currents, triggering thermoregulatory responses and complex thermal gradients throughout the body of the patient. While previous studies have implemented the Atkinson-Brezovich limit to mitigate potential harm, recent research underscores discrepancies between this threshold and clinical outcomes, necessitating a re-evaluation of this safety limit.

View Article and Find Full Text PDF

Exploring in vivo human brain metabolism at 10.5 T: Initial insights from MR spectroscopic imaging.

Neuroimage

January 2025

Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, USA. Electronic address:

Introduction: Ultra-high-field magnetic resonance (MR) systems (7 T and 9.4 T) offer the ability to probe human brain metabolism with enhanced precision. Here, we present the preliminary findings from 3D MR spectroscopic imaging (MRSI) of the human brain conducted with the world's first 10.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!