Accuracy of ultrasound to measure the cricoid cartilage diameter in children.

Acta Anaesthesiol Scand

Department of Anesthesiology, HongHui Hospital, Xi'an JiaoTong University, Xi'an, China.

Published: November 2020

Background: Ultrasonography is used to pre-operatively predict the endotracheal tube size required for intubation by measuring the cricoid cartilage diameter. This study aimed to determine the accuracy of ultrasound measurements of the transverse diameters of the cricoid cartilage in children.

Methods: We examined 50 children who underwent magnetic resonance imaging (MRI) scans at the HongHui Hospital, Xi'an Jiaotong University, from February 2019 to December 2019. Each child underwent MRI and ultrasound scans for measurement of the transverse diameters of the cricoid cartilage. The data measured by each technique were compared using Bland-Altman analysis and linear regression analysis.

Results: Results of linear regression and Bland-Altman analysis showed strong correlation in the level of agreement between MRI and ultrasound measurements (r = 0.94, P < .001). The estimated bias was 0.11 mm; precision, 0.25 mm; and the limit of agreement, -0.47 to 0.70 mm.

Conclusion: Ultrasound is a reliable tool for measuring the transverse diameters of the cricoid cartilage in children.

Download full-text PDF

Source
http://dx.doi.org/10.1111/aas.13687DOI Listing

Publication Analysis

Top Keywords

cricoid cartilage
16
accuracy ultrasound
8
cartilage diameter
8
ultrasound measurements
8
transverse diameters
8
diameters cricoid
8
mri ultrasound
8
bland-altman analysis
8
ultrasound measure
4
cricoid
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!