Objective: To provide an initial report that patients with magnetic resonance imaging (MRI) non-conditional cardiac implanted electronic device (CIED) can undergo state-of-the-art magnetic resonance imaging-guided focused (MRgFUS) ablation procedures with careful planning and integration of the procedure into an established CIED MRI practice.
Patient And Methods: We describe an MRgFUS ablation treatment of lumbar facet joints in a patient with an MRI non-conditional CIED (pacemaker), completed in accordance with our institutional CIED/MRI practice guidelines.
Results: A risk-benefit analysis by a coordinated multidisciplinary team before this treatment was performed to account for the risks associated with the MRI non-conditional pacemaker in the context of the MRgFUS procedure.
Conclusion: The patient had no adverse cardiac event during or following this procedure.