Objectives: To investigate via computer simulation, an optimised tube voltage (kVp) range for caesium iodide (CsI)-based digital radiography (DR) of the abdomen, pelvis and lumbar spine.

Methods: Software capable of simulating abdomen, pelvis and spine radiographs was used. Five evaluators graded clinical image criteria in images of 20 patients at tube voltages ranging from 60 to 120 kVp in 10 kVp increments. These criteria were scored blindly against the same patient reconstructed at a specific reference kVp. Linear mixed effects analysis was used to evaluate image scores for each criterion and test for statistical significance.

Results: Score was dependent on tube voltage and image criteria; both were statistically significant. All criteria for all anatomies scored very poorly at 60 kVp. Scores for abdomen, pelvis and spine imaging peaked at 70, 70 and 100 kVp, respectively, but other kVp values were not significantly poorer.

Conclusions: Results indicate optimum tube voltages of 70 kVp for abdomen and pelvis (with an optimum range 70-120 kVp), and 100 kVp (optimum range 80-120 kVp) for lumbar spine.

Advances In Knowledge: There are no recommendations for optimised tube voltage parameters for DR abdomen, pelvis or lumbar spine imaging. This study has investigated and recommended an optimal tube voltage range.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7548356PMC
http://dx.doi.org/10.1259/bjr.20200565DOI Listing

Publication Analysis

Top Keywords

abdomen pelvis
24
tube voltage
20
kvp
12
pelvis spine
12
spine imaging
12
voltage range
8
kvp abdomen
8
digital radiography
8
optimised tube
8
pelvis lumbar
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!