This paper proposes a new method to reject pile-up pulse continuous zone. This paper first simulates the differences between three kinds trapezoidal shaping methods of which are trapezoidal shaping without width discrimination, with fixed width discrimination or with changing width discrimination, and then carry out experiment with silicon drift detector (SDD) and Fe source in laboratory, and the result of the experiment is that the ratio of peak to background (P/B) increases while the characteristic peaks in high energy section are not affected. Result of the simulation and experiment shows that the method proposed in this paper makes progress in pile-up pulse continuous zone rejection under the circumstance of high-count rates.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.apradiso.2020.109319 | DOI Listing |