Instrumented triple single-leg hop test: A validated method for ambulatory measurement of ankle and knee angles using inertial sensors.

Clin Biomech (Bristol)

Department of Mechanical Engineering, University of Alberta, Donadeo Innovation Centre for Engineering, Edmonton, Alberta T6G 1H9, Canada. Electronic address:

Published: December 2020

Background: Hop tests are commonly used in clinical environments to measure function after sport-related knee injuries. Joint angle measurement during hopping is feasible in research-based environments equipped with motion-capture systems. Employing these systems in clinical research settings is inefficient, given the associated cost, preparation time, and expertise required to administer and interpret the findings. Therefore, this study aimed to introduce a wearable system comprising three inertial measurement units for 3D joint angular measurement during horizontal hop tests, validate the joint angles against a camera-based system, and evaluate its applicability in clinical research environments.

Methods: Ten able-bodied participants were outfitted with three inertial measurement units during triple single-leg hop trials. 3D knee and ankle angles were calculated using the strap-down integration method, and results were compared with camera-based joint angles. Additionally, knee and ankle range of motions (RoMs) during bilateral triple single-leg hop trials were compared for 22 participants with unilateral sport-related knee injuries and 10 uninjured participants.

Findings: Estimated angles had root-mean-square and RoM error medians of less than 2.3 and 3.2 degrees for both joints, and correlation coefficients of above 0.92 when compared with the camera-based system, for all hop phases. Injured participants had smaller sagittal ankle RoM (P = .008) on their injured side, during the third hop. Concurrently, they demonstrated smaller knee RoM symmetry indices (P = .017) and injured knee sagittal RoMs (P = .009) compared to uninjured participants.

Interpretation: The introduced system had appropriate accuracy to highlight post-injury modifications in hopping kinematics and reveal noteworthy differences in RoM of clinical samples.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiomech.2020.105134DOI Listing

Publication Analysis

Top Keywords

triple single-leg
12
single-leg hop
12
hop tests
8
sport-related knee
8
knee injuries
8
three inertial
8
inertial measurement
8
measurement units
8
joint angles
8
camera-based system
8