Monte Carlo-based scatter correction for the SMARTZOOM collimator.

EJNMMI Phys

Radiology and Nuclear Medicine, University Medical Center Utrecht, Utrecht University, P.O. Box 85500, 3508, GA, Utrecht, The Netherlands.

Published: July 2020

Background: Myocardial perfusion imaging is a commonly performed SPECT protocol and hence it would be beneficial if its scan duration could be shortened. For traditional gamma cameras, two developments have separately shown to allow for a shortened scan duration: (i) reconstructing with Monte Carlo-based scatter correction instead of dual-energy window scatter correction and (ii) acquiring projections with the SMARTZOOM collimator instead of a parallel-hole collimator. This study investigates which reduction in scan duration can be achieved when both methods are combined in a single system.

Results: The SMARTZOOM collimator was implemented in a Monte Carlo-based reconstruction package and the implementation was validated through image quality phantom experiments. The potential for scan duration reduction was evaluated with a phantom configuration that is realistic for myocardial perfusion imaging. The original reconstruction quality was achieved in 76 ± 8% of the original scan duration when switching from dual-energy window scatter correction to Monte Carlo-based scatter correction. The original reconstruction quality was achieved in 56 ± 13% of the original scan duration when switching from the parallel-hole to the SMARTZOOM collimator. After combining both methods in a single system, the original reconstruction quality was achieved in 34 ± 7% of the original scan duration.

Conclusions: Monte Carlo-based scatter correction combined with the SMARTZOOM collimator can further decrease the scan duration in myocardial perfusion imaging.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7376766PMC
http://dx.doi.org/10.1186/s40658-020-00318-7DOI Listing

Publication Analysis

Top Keywords

scan duration
28
scatter correction
24
monte carlo-based
20
smartzoom collimator
20
carlo-based scatter
16
myocardial perfusion
12
perfusion imaging
12
original reconstruction
12
reconstruction quality
12
quality achieved
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!