  • The study aimed to evaluate the primary stability of a press-fit cup in a model of acetabular defects filled with either bone chips or two different bone graft substitutes.
  • Three test groups were used, consisting of cancellous bone chips, tricalcium phosphate tetrapods with a collagen matrix, and bioactive glass combined with a polyethylene glycol-glycerol matrix.
  • Results showed that while bone chips exhibited the greatest displacements, a combination of bioactive glass and the PEG matrix had the least displacement and showed potential as a viable alternative to bone chips, although further research on different defect types is needed.

Article Abstract

The objectives of this study were to (a) assess primary stability of a press-fit cup in a simplified acetabular defect model, filled with compacted cancellous bone chips, and (b) to compare the results with primary stability of a press-fit cup combined with two different types of bone graft substitute in the same defect model. A previously developed acetabular test model made of polyurethane foam was used, in which a mainly medial contained defect was implemented. Three test groups (N = 6 each) were prepared: Cancellous bone chips (bone chips), tricalciumphosphate tetrapods + collagen matrix (tetrapods + coll), bioactive glass S53P4 + polyethylene glycol-glycerol matrix (b.a.glass + PEG). Each material was compacted into the acetabulum and a press-fit cup was implanted. The specimens were loaded dynamically in the direction of the maximum resultant force during level walking. Relative motion between cup and test model was assessed with an optical measurement system. At the last load step (3000 N), inducible displacement was highest for bone chips with median [25th percentile; 75th percentile] value of 113 [110; 114] µm and lowest for b.a.glass + PEG with 91 [89; 93] µm. Migration at this load step was highest for b.a.glass + PEG with 868 [845; 936] µm and lowest for tetrapods + coll with 491 [487; 497] µm. The results show a comparable behavior under load of tetrapods + coll and bone chips and suggest that tetrapods + coll could be an attractive alternative to bone chips. However, so far, this was found for one specific defect type and primary stability should be further investigated in additional/more severe defects.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jor.24810DOI Listing

Publication Analysis

Top Keywords

bone chips
24
primary stability
16
press-fit cup
16
stability press-fit
12
defect model
12
acetabular defect
8
cancellous bone
8
test model
8
load step
8
bone
7

Similar Publications

The objective of this study was to tailor an osteoinductive scaffold for alveolar bone regeneration and around immediately placed implants in extraction sockets of dogs. Tailored amorphous multiporous bioactive glass (TAMP -BG) was prepared and characterized for bioactivity and response of human alveolar bone marrow mesenchymal stem cells (hABMSCs). Extraction sockets of twenty-two male mongrel dogs received TAMP-BG in the right side around implant in the distal socket of the mandibular fourth premolar (P4), while the adjacent empty mesial socket of the same tooth was filled with the same graft.

View Article and Find Full Text PDF

Unrelated bone marrow transplantation (BMT) is a curative treatment for hematological malignancies. While HLA mismatch is a recognized risk factor in unrelated BMT, the significance of non-HLA single nucleotide polymorphisms (SNPs) remains uncertain. Cytokines play key roles in several aspects of unrelated BMT.

View Article and Find Full Text PDF

Craniometaphyseal dysplasia (CMD), a rare craniotubular disorder, occurs in an autosomal dominant (AD) or autosomal recessive (AR) form. CMD is characterized by hyperostosis of craniofacial bones and metaphyseal flaring of long bones. Many patients with CMD suffer from neurological symptoms.

View Article and Find Full Text PDF

Monitoring and optimization of the microenvironment in a gravity-driven microfluidic system placed on a slow-tilting table.

J Biosci Bioeng

January 2025

Department of Bioengineering, Nagaoka University of Technology, 1603-1 Kamitomioka-machi, Nagaoka, Niigata 940-2188, Japan; Department of Science of Technology Innovation, Nagaoka University of Technology, 1603-1, Kamitomioka-machi, Nagaoka, Niigata, 940-2188 Japan. Electronic address:

Gravity-driven microfluidic chips offer portability and flexibility in different settings because pumps and connecting tubes are unnecessary for driving fluid flow. In a previous study, human induced pluripotent stem cells were cultured using gravity-driven microfluidics, with the liquid flow rate regulated by a tilting table. However, instability in cell culture has been observed, occasionally leading to cell death owing to unknown causes.

View Article and Find Full Text PDF

Sandblasting and acid etching are common procedures used to treat implant surfaces, enhancing osseointegration and improving clinical success rates. This clinical study aimed to evaluate the long-term outcomes of sandblasted and acid-etched implants. A total of 303 implants were placed in 114 partially and totally edentulous patients using a two-stage surgical technique and an early loading protocol (6-8 weeks).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!