Purpose: To characterize the Poisson effect in response to angioplasty and stent placement in veins and identify potential implications for guiding future venous-specific device design.
Materials And Methods: In vivo angioplasty and stent placement were performed in 3 adult swine by using an established venous stenosis model. Iron particle endothelium labeling was performed for real-time fluoroscopic tracking of the vessel wall during intervention. A finite-element computational model of a vessel was created with ADINA software (version 9.5) with arterial and venous biomechanical properties obtained from the literature to compare the response to radial expansion.
Results: In vivo angioplasty and stent placement in a venous stenosis animal model with iron particle endothelium labeling demonstrated longitudinal foreshortening that correlated with distance from the center of the balloon (R = 0.87) as well as adjacent segment narrowing that correlated with the increase in diameter of the treated stenotic segment (R = 0.89). Finite-element computational analysis demonstrated increased Poisson effect in veins relative to arteries (linear regression coefficient slope comparison, arterial slope 0.033, R = 0.9789; venous slope 0.204, R = 0.9975; P < .0001) as a result of greater longitudinal Young modulus in veins compared with arteries.
Conclusions: Clinically observed adjacent segment narrowing during venous angioplasty and stent placement is a result of the Poisson effect, with redistribution of radially applied force to the longitudinal direction. The Poisson effect is increased in veins relative to arteries as a result of unique venous biomechanical properties, which may be relevant to consider in the design of future venous interventional devices.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jvir.2020.02.033 | DOI Listing |
Cureus
November 2024
Division of Interventional Radiology, Columbia University Irving Medical Center, New York, USA.
Aim This study aims to evaluate the safety and efficacy of transjugular intrahepatic portosystemic shunt (TIPS) check and revision procedures performed in a freestanding interventional radiology (IR) outpatient facility. Methodology A total of 40 patients (male 31:female 9, median age 60 years old) underwent a TIPS check and/or revision at a freestanding IR outpatient facility between 2009 and 2017. Procedures were performed using a mobile C-arm unit under intravenous (IV) moderate sedation, with the patient discharged home on the same day.
View Article and Find Full Text PDFActa Radiol
December 2024
Department of Interventional Radiology, The First Affiliated Hospital of Nanjing Medical University, Nanjing, PR China.
Background: In-stent restenosis (ISR) is a potential severe complication that occurs in patients with severe carotid artery narrowing after carotid angioplasty and stent placement. However, this phenomenon has not been fully studied in the context of interventional treatment for chronic internal carotid artery occlusion (CICAO).
Purpose: To quantify the ISR rate and identify the risk factors leading to this event.
J Int Med Res
December 2024
Department of Cardiology, General Hospital of Ningxia Medical University, Yinchuan, People's Republic of China.
Objective: This study aimed to compare the efficacy and safety of intravascular ultrasound (IVUS)-guided coronary intravascular lithotripsy and rotational atherectomy in treating severe coronary artery calcification.
Methods: A retrospective analysis of 60 patients who underwent percutaneous coronary intervention at the General Hospital of Ningxia Medical University from October 2022 to August 2023 was conducted. The patients were divided into two groups: 30 received IVUS-guided coronary intravascular lithotripsy and 30 underwent IVUS-guided rotational atherectomy.
Front Cardiovasc Med
December 2024
Department of Nephrology, Guangdong Provincial People's Hospital (Guangdong Academy of Medical Sciences), Southern Medical University, Guangzhou, China.
Background: The number of people undergoing maintenance hemodialysis is increasing rapidly worldwide. Central vein stenosis (CVS) is a common vascular complication in undergoing hemodialysis, especially those with a history of catheterization. This study aimed to investigate the characteristics of CVS and the clinical effectiveness of percutaneous transluminal angioplasty (PTA) alone and sequential percutaneous transluminal stenting (PTS) in hemodialysis patients with CVS.
View Article and Find Full Text PDFJAMA Cardiol
December 2024
The Zena and Michael A. Wiener Cardiovascular Institute, Icahn School of Medicine at Mount Sinai, New York, New York.
Importance: Drug-coated balloon (DCB) angioplasty has emerged as an alternative to drug-eluting stent (DES) implantation for percutaneous coronary intervention (PCI) in patients with coronary in-stent restenosis (ISR) as well as de novo coronary artery disease.
Observations: DCBs are balloons coated with antiproliferative agents and excipients, whose aim is to foster favorable vessel healing after appropriate lesion preparation. By providing homogeneous antiproliferative drug delivery in the absence of permanent foreign body implantation, DCBs offer multiple advantages over DES, including preservation of vessel anatomy and function and positive vessel remodeling.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!