The aim of this study is to compare the accuracy of Airgo™, a non-invasive wearable device that records breath, with respect to a gold standard. In 21 healthy subjects (10 males, 11 females), four parameters were recorded for four min at rest and in different positions simultaneously by Airgo™ and SensorMedics 2900 metabolic cart. Then, a cardio-pulmonary exercise test was performed using the Erg 800S cycle ergometer in order to test Airgo™'s accuracy during physical effort. The results reveal that the relative error median percentage of respiratory rate was of 0% for all positions at rest and for different exercise intensities, with interquartile ranges between 3.5 (standing position) and 22.4 (low-intensity exercise) breaths per minute. During exercise, normalized amplitude and ventilation relative error medians highlighted the presence of an error proportional to the volume to be estimated. For increasing intensity levels of exercise, Airgo™'s estimate tended to underestimate the values of the gold standard instrument. In conclusion, the Airgo™ device provides good accuracy and precision in the estimate of respiratory rate (especially at rest), an acceptable estimate of tidal volume and minute ventilation at rest and an underestimation for increasing volumes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7412454PMC
http://dx.doi.org/10.3390/s20143943DOI Listing

Publication Analysis

Top Keywords

airgo™ device
8
metabolic cart
8
rest exercise
8
gold standard
8
relative error
8
respiratory rate
8
exercise
6
rest
5
comparison airgo™
4
device metabolic
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!