Purpose: We aimed to evaluate and compare the shapeability and stability of five microcatheters commonly used in interventional radiology after steam shaping and manual shaping.
Methods: Steam shaping was performed using three mandrels of different angles: L(S) shape (90°), U(S) shape (180°), and O(S) shape (360°). Three manual shapes-L(M), U(M), and O(M)-were made to have a similar angle to their steam-shaped counterparts. The stability of the microcatheters was evaluated by passing them through a 5 F catheter and inserting microguidewires. The tip angles of the microcatheters and the angle change rates were compared between groups.
Results: The mean angle of the microcatheters after steam shaping was 42.4°-54.1° for L(S) shape, 80.2°-96.7° for U(S) shape, and 130.7°-150.8° for O(S) shape. Five microcatheters showed significantly different mean angle reductions after passing through the 5 F catheter (17.4%-30.3%) and inserting microguidewires (24.1%-61.2%). Different microguidewires also caused significantly different mean angle reductions (34.6%-50.8%). The reduced angle caused by the guidewire was almost completely recovered after withdrawing it (93.2%-101.6%). Although manual-shaped microcatheters showed a 4.2%-6.3% greater angle reduction than steam-shaped microcatheters after passing through the 5 F catheter, the final tip angle was not significantly different between the two groups and was within 10%.
Conclusion: The tip angle of the microcatheters after steam shaping using mandrels may differ depending on the shape of the mandrel and the type of microcatheter used, and the stability varies depending on the type of microcatheter. The manual shaping of microcatheters can be a good alternative to steam shaping.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7490016 | PMC |
http://dx.doi.org/10.5152/dir.2020.19314 | DOI Listing |
J Prosthodont Res
January 2025
Department of Orthodontics, Osaka Dental University, Hirakata, Japan.
Purpose: To perform vertical bone augmentation on rat parietal bone by coating the inner surface of dense polytetrafluoroethylene (d-PTFE) domes with hydroxyapatite (HA) using Erbium Yttrium Aluminum Garnet (Er:YAG) pulsed laser deposition in a rat model.
Methods: The d-PTFE plate surface, α-tricalcium phosphate (α-TCP) coating, and HA coating were measured using scanning electron microscopy and X-ray diffraction to confirm the replacement of α-TCP with HA via high-pressure steam sterilization. The dome was glued to the center of the rat parietal bone and closed with periosteal and epithelial sutures.
Mar Pollut Bull
December 2024
Biological Oceanography Division, CSIR-National Institute of Oceanography, Dona Paula, Panaji, Goa 403004, India.
Microplastics (MPs, <5 mm) are widespread in coastal ecosystems and pose a growing global concern; however, their presence in deep-sea environments remains underexplored, especially in the Indian region. This study addresses this gap by providing the first comprehensive documentation of MPs in the Central Indian Ocean Basin (CIOB) at a depth of 5000 m, marking the initial effort to assess their presence and abundance in deep-sea core samples. The study investigated the MP concentration, composition and potential sources, revealing a size range between 10 μm and 4900 μm, with average abundances recorded at BC20 (10.
View Article and Find Full Text PDFFoods
December 2024
Key Laboratory of Tea Resources Comprehensive Utilization (Co-Construction by Ministry and Province), Ministry of Agriculture and Rural Affairs, Fruit and Tea Research Institute, Hubei Academy of Agricultural Sciences, Wuhan 430064, China.
Volatile constituents are critical to the flavor of tea, but the changes in Enshi Yulu tea during the processing have not been clearly understood. Using headspace solid phase microextraction combined with gas chromatography-mass spectrometry (HS-SPME/GC-MS) techniques, we analyze the aroma components of Enshi Yulu tea and changes in them during the processing stages. In total, 242 volatile compounds were identified.
View Article and Find Full Text PDFClin Neuroradiol
November 2024
Department of Neuroradiology, King's College Hospital NHS Foundation Trust, London, UK.
Eur J Radiol
December 2024
Neurovascular Center, Changhai Hospital, Naval Medical University, Shanghai, China. Electronic address:
Objective: To compare the coating damage and shape retention quality of microcatheter by steam (ST), hot air gun (HA), and HA + ST shaping.
Method: SL-10, XT-17, and Q-track microcatheters were included in the study. MarSurf CM explorer (a non-contact 3D confocal microscopy system) and MATLAB software were used to detect the surface coating damage.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!