Feasibility of customizing titanium implant with three-dimensional CT imaging of low dose in skull.

Medicine (Baltimore)

Department of Radiology, Shaoxing People's Hospital, Shaoxing Hospital of Zhejiang University, Shaoxing, Zhejiang Province, China.

Published: July 2020

Object: To explore the feasibility and practicability of making virtual three-dimensional model of skull defect and customizing titanium implant by skull three-dimensional CT examination of low dose.

Methods: Sixty patients with skull defects who underwent skull three-dimensional CT before cranioplasty were randomly divided into 4 groups: group A (conventional dose 120 peak Kilovoltage (kVp), 150 tube current time product (mAs)), low dose group B (120 kVp, 50 mAs), low dose group C (100 kVp, 50 mAs), low dose group D (100 kVp, 30 mAs). After the scanning, we compared radiation doses and image quality among the groups. The CT data were sent to the reconstruction company to produce accurate titanium implants, and neurosurgeons performed cranioplasty. After the operation, patients immediately underwent head CT scans to confirm the accuracy of the implantation position, and a series of clinical functions were evaluated.

Results: There were significant differences in dose length product (DLP) and effective dose (ED) among the 4 groups (P < .001). The volume CT dose index (CDTIvol), DLP, and ED in group D were, respectively, 87.1%, 86.9%, and 87.3% lower than those in group A (P < .001). All images quality were at or above the general level, and there was no statistical difference (P > .05). Titanium implants were successfully manufactured, every cranioplasty was carried out smoothly, and the clinical function of patients recovered well.

Conclusion: Customizing titanium implant with three-dimensional CT imaging of low dose in skull not only met the clinical requirements, but also significantly reduced the radiation dose and hazard.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7360305PMC
http://dx.doi.org/10.1097/MD.0000000000021009DOI Listing

Publication Analysis

Top Keywords

low dose
16
mas low
12
dose group
12
kvp mas
12
customizing titanium
8
titanium implant
8
skull three-dimensional
8
group 100
8
100 kvp
8
dose
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!