Objective: The [Formula: see text]-Ray Evaluation with Iodoamphetamine for Cerebral blood flow Assessment (REICA) method is a new method for cerebral blood flow (CBF) quantification with SPECT and [I] IMP, which does not require invasive arterial blood sampling, or the use of a regression formula. Moreover, it takes into consideration tracer retention, while the Graph-Plot (GP) method does not. This study aims to confirm the reliability of the REICA method in calculating CBF by accounting for pulmonary tracer retention artefact.

Methods: Data of 100 patients were retrospectively analyzed. Washout rate (WR) was defined as 1- (max count /min count) in the lungs. The bias due to WR was analyzed by dividing the data into high/low WR groups and a multiple regression analysis was also performed.

Results: The REICA method did not show significant bias according to the [Formula: see text] test ([Formula: see text]), while the GP method showed significant bias ([Formula: see text]). Both the REICA and GP methods showed significant correlation with autoradiography (ARG) method ([Formula: see text], [Formula: see text]) and WR ([Formula: see text], [Formula: see text]) when analyzed via multiple regression analysis with covariates of age and sex.

Conclusions: The REICA method is a reliable method for CBF quantification, which is unaffected by tracer washout through the lungs unlike the GP method.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12149-020-01499-yDOI Listing

Publication Analysis

Top Keywords

[formula text]
28
reica method
20
cerebral blood
12
blood flow
12
text] [formula
12
method
11
[i] imp
8
graph-plot method
8
[formula
8
cbf quantification
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!