Background: A suitable fundus camera for telemedicine screening can expand the scale of eye care service. The purpose of this study was to compare a handheld nonmydriatic digital fundus camera and a conventional mydriatic fundus camera according to the image quality of their photographs and usability of those photographs to accurately diagnose various retinal diseases.
Methods: A handheld nonmydriatic fundus camera and conventional fundus camera were used to take fundus photographs of outpatients at an ophthalmic clinic before and after pupillary dilation. Image quality and diagnostic agreement of the photos were graded by two masked and experienced retinal specialists.
Results: A total of 867 photographs of 393 eyes of 200 patients were collected. Approximately 80% of photos taken under nonmydriasis status using the handheld nonmydriatic fundus camera had good (55.7%) or excellent (22.7%) image quality. The overall agreement of diagnoses between the doctors was more than 90%. When the handheld nonmydriatic fundus camera was used after mydriasis, the proportion of images with good (45%) or excellent (49.7%) quality reached 94.7% and diagnostic agreement was 93.4%. Lens opacity was associated with the quality of images obtained using the handheld camera (p = 0.041), and diagnosis disagreement for handheld camera images was associated with preexisting diabetes diagnosis (p = 0.009). Approximately 40% of patients expressed preference for use of the handheld nonmydriatic camera.
Conclusion: This study demonstrated the effectiveness of the handheld nonmydriatic fundus camera in clinical practice and its feasibility for telemedicine screening of retinal diseases.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7526587 | PMC |
http://dx.doi.org/10.1097/JCMA.0000000000000382 | DOI Listing |
Heliyon
July 2024
D-Eye Srl, Padova, 35131, Italy.
Widespread screening is crucial for the early diagnosis and treatment of glaucoma, the leading cause of visual impairment and blindness. The development of portable technologies, such as smartphone-based ophthalmoscopes, able to image the optical nerve head, represents a resource for large-scale glaucoma screening. Indeed, they consist of an optical device attached to a common smartphone, making the overall device cheap and easy to use.
View Article and Find Full Text PDFBiomed Opt Express
January 2025
Department of Biomedical Engineering, University of Illinois Chicago, Chicago, IL 60607, USA.
The choroid, a critical vascular layer beneath the retina, is essential for maintaining retinal function and monitoring chorioretinal disorders. Existing imaging methods, such as indocyanine green angiography (ICGA) and optical coherence tomography (OCT), face significant limitations, including contrast agent requirements, restricted field of view (FOV), and high costs, limiting accessibility. To address these challenges, we developed a nonmydriatic, contrast agent-free fundus camera utilizing transcranial near-infrared (NIR) illumination.
View Article and Find Full Text PDFCJEM
January 2025
Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, ON, Canada.
Objectives: Fundoscopy is crucial in the emergency department to identify or rule out serious ocular and neurological conditions. Despite its clinical importance, fundoscopy is often omitted due to the technical challenges associated with traditional direct ophthalmoscopy, particularly for non-ophthalmologists. This study examines emergency physicians' practices, confidence levels, and training related to various modalities of fundoscopy including traditional direct ophthalmoscopes, binocular indirect ophthalmoscopes, panoptic ophthalmoscopes, slit lamp fundoscopy and fundus cameras; and explores the potential role of alternative modalities, such as fundus cameras, in Canadian emergency departments.
View Article and Find Full Text PDFPol J Vet Sci
June 2024
Department and Clinic of Animal Surgery, Faculty of Veterinary Medicine, University of Life Sciences in Lublin, Głęboka 30, 20-612 Lublin, Poland.
The aim of the study was to determine the thickness of choroidal layers in mixed breed dogs suffering from retinal atrophy (RA) and showing symptoms of progressive retinal atrophy (PRA), with the use of SD-OCT. The study was performed on 50 dogs divided into two groups: 25 dogs diagnosed with retinal atrophy (RA) with PRA symptoms aged 1.5-14 years and 25 healthy dogs aged 2-12 years.
View Article and Find Full Text PDFNPJ Digit Med
December 2024
Department of Ophthalmology and Visual Sciences, University of Iowa, Iowa City, IA, USA.
Where adopted, Autonomous artificial Intelligence (AI) for Diabetic Retinal Disease (DRD) resolves longstanding racial, ethnic, and socioeconomic disparities, but AI adoption bias persists. This preregistered trial determined sensitivity and specificity of a previously FDA authorized AI, improved to compensate for lower contrast and smaller imaged area of a widely adopted, lower cost, handheld fundus camera (RetinaVue700, Baxter Healthcare, Deerfield, IL) to identify DRD in participants with diabetes without known DRD, in primary care. In 626 participants (1252 eyes) 50.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!