A multicolor fluorescence imaging device was recently developed for image-guided surgery. However, conventional systems are typically bulky and function with two cameras. To overcome these issues, we developed an economical home-built fluorescence imaging device based on a single RGB-IR sensor that can acquire both color and fluorescence images simultaneously. The technical feasibility of RGB-IR imaging was verified in chicken breast tissue using fluorescein isothiocyanate (FITC), cyanine 5 (Cy5), and indocyanine green (ICG) as fluorescent agents. The minimum sensitivities for FITC, Cy5, and ICG were 0.200 µM, 0.130 µM, and 0.065 µM, respectively. In addition, we validated the fluorescence imaging of this device during a minimally invasive procedure using smURFP-labeled probiotics, which emit a spectrum similar to that of Cy5. Our preliminary study of the tissue suggests that Cy5 and ICG are good candidates for deep tissue imaging. In addition, the tumor-specific amplification process was visualized using cancer cells incubated with probiotics that had been labeled with a fluorescent protein. Our approach indicates the potential for screening of tumors in rodent tumor models.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7316003 | PMC |
http://dx.doi.org/10.1364/BOE.391417 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!