A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Evaluation of the Aptima™ HBV Quant Dx assay for semi-quantitative HBV viral load from dried blood spots. | LitMetric

Background: The detection and quantification of hepatitis B virus (HBV) DNA from dried blood spots (DBS) is a major tool for chronic hepatitis B management in resource-limited settings. This strategy fits in perfectly with the hepatitis control plan promoted by the World Health Organization. However, few commercial methods are validated for viral load (VL) measurement on DBS.

Objectives: Our objective was to evaluate the performance of the HBV VL measurement of the Aptima™ HBV Quant Dx assay on DBS compared to plasma samples on the Panther® platform (Hologic).

Study Design: 266 whole blood samples for routine measurement were included. Five spots of 75 μL of whole blood were loaded onto a card before centrifugation and plasma settling.

Results: 149 samples were quantifiable and 117 were not detected. We achieved excellent linearity (r² = 0.994) over a wide range of measurements suitable for clinical practice, and a 95 % lower limit of detection (LLOD-95 %) at 2.65 log IU/mL (445 IU/mL). A good performance of this assay was observed for samples with HBV VL > LLOD-95 % and 100 % of samples were detected if HBV VL was above 2.95 log IU/mL. The correlation between the two matrices for quantitative VLs was good (r² = 0.978) with a very low bias (-0.002 log IU/mL).

Conclusion: The Aptima™ assay can properly detect and quantify HBV DNA in DBS, providing a satisfactory use in clinical monitoring and therapeutic decisions. DBS represents an excellent alternative to plasma, especially in resource-limited countries, while maintaining the performance and advantages of an automated technique.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcv.2020.104524DOI Listing

Publication Analysis

Top Keywords

hbv
8
aptima™ hbv
8
hbv quant
8
quant assay
8
viral load
8
dried blood
8
blood spots
8
hbv dna
8
log iu/ml
8
samples
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!