Background: Management of proximal humerus fractures is challenging, especially in elderly. Locking plating is a common surgical treatment option. The Proximal Humerus Internal Locking System (plate-A) has shown to lower complication rates compared to conventional plates, but is associated with impingement risk, which could be avoided using Peri-articular Proximal Humerus Plate (plate-B). Nevertheless, biomechanical performance and optimal screw configuration of plate-B is unknown. The aim of this study was to evaluate different screw configurations of plate-B and compare with plate-A using finite element analyses.

Methods: Twenty-six proximal humerus models were osteotomised to create unstable three-part fractures, fixed with either of the two plates, and tested under three anatomical loading conditions using a previous established and validated finite element simulation framework. Various clinically relevant screw configurations were investigated for both plates and compared based on the predicted peri-implant bone strain, being a validated surrogate of cyclic cut-out failure.

Findings: Besides increasing the number of screws, the placement of the posterior screws in combination with the calcar screw in the plate-B significantly decreased the predicted failure risk. Generally, plate-A had a lower predicted failure risk than plate-B.

Interpretation: The posterior and calcar screws may be prioritized in plate-B. Compared to plate-A, the more distal positioning, less purchase in the posterior aspect and a smaller screw spread due to not fitting of the most distal calcar screw in most investigated subjects led to a significantly higher predicted failure risk for most plate-B configurations. The findings of the simulations study require clinical corroboration.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiomech.2020.105097DOI Listing

Publication Analysis

Top Keywords

proximal humerus
20
screw configurations
12
finite element
12
predicted failure
12
failure risk
12
optimal screw
8
plate-a lower
8
calcar screw
8
screw
7
plate-b
6

Similar Publications

Background: Proximal humerus fractures (PHF) are common with approximately 30% requiring surgical intervention. This ranges from open reduction internal fixation (ORIF) to shoulder arthroplasty (including hemiarthroplasty, total shoulder arthroplasty (TSA), reverse total shoulder arthroplasty (RTSA)). The aim of this study was to assess trends in operative interventions for PHF in an Australian population.

View Article and Find Full Text PDF

Objective: This study evaluated the effect of three-dimensional (3D) volumetric humeral canal fill ratios (VFR) of reverse shoulder arthroplasty (RSA) short and standard stems on biomechanical stability and bone deformations in the proximal humerus.

Methods: Forty cadaveric shoulder specimens were analyzed in a clinical computed tomography (CT) scanner allowing for segmentation of the humeral canal to calculate volumetric measures which were verified postoperatively with plain radiographs. Virtual implant positioning allowed for group assignment (VFR < 0.

View Article and Find Full Text PDF

Shoulder injuries, specifically proximal humeral fractures, are uncommon in skeletally immature patients. The anatomic characteristics of the humerus are determined by its ossification development, which is quite particular in the proximal segment where four principal segments have been outlined as fracture components in adults. Hereby, we present the case of an 18-year-old female who suffered a traffic accident that elicited a proximal humeral injury through the physeal line of the anatomical neck, as well as the treatment given and her clinical outcome with an effort to highlight the complexity of the diagnosis and management of this injury given the transitional bone age.

View Article and Find Full Text PDF

Biomechanical study of two different fixation methods for the treatment of Neer III proximal humerus fractures.

BMC Musculoskelet Disord

December 2024

Department of Orthopedic Surgery, Shanghai Sixth People's Hospital, Shanghai Jiao Tong University School of Medicine, Jinshan Branch, Jinshan District Central Hospital, Shanghai, 201500, China.

Background: The lateral locking plate for the proximal humerus is currently the most commonly used surgical procedure for the treatment of elderly proximal humeral comminuted fractures. Previous studies have found that the rate of postoperative complications in patients of proximal humerus fractures with medial column involvement is relatively high. Through biomechanical methods, this study aims to investigate the effectiveness of the conventional lateral locking plate fixation along with the addition of the metacarpal supporting plate on the medial column in the treatment for proximal humeral fractures involving the medial column.

View Article and Find Full Text PDF

Background: The optimal treatment of 3- and 4-part proximal humeral fractures in older adults remains controversial. This aim of this study was compare patient reported outcomes following reverse shoulder arthroplasty (RSA) or non-operative management in patients over 60 years old.

Methods: A retrospective review was undertaken of patients following 3- or 4-part proximal humeral fractures treated with RSA or non-operative treatment with minimum 2-year follow-up.

View Article and Find Full Text PDF