Purpose: Radiation exposure during coronary angiography is potentially harmful to patients and operators. However, there are limited data on the effects of a low-dose radiation angiography. We evaluated the feasibility and effectiveness of a reduced radiation dose protocol during invasive coronary angiography.

Methods: One hundred three consecutive patients who underwent coronary angiography were enrolled and randomized to low- or conventional dose protocols (LDP versus CDP). The LDP consists of 10 frames per second during fluoroscopy and half the radiation dose of CDP during cineangiography. Image quality was assessed using a Likert rating scale by an independent radiologist. The radiation dose was estimated with dose-area product (DAP) and air-kerma (AK).

Results: Body weight and waist circumference are well correlated with the level of DAP and AK. Exposure time and total images and frame counts in cineangiography were similar in both groups. There was a marked reduction of the estimated radiation dose (DAP and AK) in the LDP group compared to the CDP group without significant compromise in image quality (total DAP: LDP 1980.1 ± 1163.7 vs. CDP 3434.2 ± 2188.1 μGym p = 0.001; total AK: 279.6 ± 159.3 vs. 493.8 ± 280.6 mGy, p < 0.001).

Conclusion: The LDP reduced the total estimated radiation dose compared to the CDP without a significant loss of diagnostic information. A LDP may be a viable strategy to protect patients and medical staff from the hazards of radiation in the cardiac catheterization laboratory.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2020.109120DOI Listing

Publication Analysis

Top Keywords

radiation dose
20
estimated radiation
12
coronary angiography
12
radiation
9
radiation exposure
8
conventional dose
8
dose protocol
8
protocol invasive
8
invasive coronary
8
image quality
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!