Objectives: To evaluate the performance of Influ A + B K-SeT®, a rapid influenza antigen detection kit, in order to determine whether its implementation as a point-of-care test in the emergency room (ER) is justified.
Methods: Nasopharyngeal swabs (n = 170) were collected in triplicates from patients with suspected influenza infection. One sample was tested by the Influ A + B K-SeT® kit and the other two samples were analyzed with Simplexa™ Flu A/B & RSV Direct Kit and Xpert® Flu, two RT-PCR-based assays routinely used in the ER.
Results: The sensitivity and specificity of the Influ A + B K-SeT® kit were 91.2%-94.5% and 98.2%, respectively, compared to the Xpert® Flu assay and the Simplexa™ Flu A/B & RSV Direct Kit, respectively.
Conclusions: Considering the satisfactory diagnostic performance of the Influ A + B K-SeT® test, its implementation as a point-of-care is expected to aid in decision-making, minimize time-to-result and reduce healthcare costs.