Performance of the Influ a + B K-SeT® assay as compared to two RT-PCR assays for detection of influenza virus.

Diagn Microbiol Infect Dis

Clinical Microbiology Laboratory, Baruch Padeh Medical Center, Poriya, 15208, Israel. Electronic address:

Published: September 2020

Objectives: To evaluate the performance of Influ A + B K-SeT®, a rapid influenza antigen detection kit, in order to determine whether its implementation as a point-of-care test in the emergency room (ER) is justified.

Methods: Nasopharyngeal swabs (n = 170) were collected in triplicates from patients with suspected influenza infection. One sample was tested by the Influ A + B K-SeT® kit and the other two samples were analyzed with Simplexa™ Flu A/B & RSV Direct Kit and Xpert® Flu, two RT-PCR-based assays routinely used in the ER.

Results: The sensitivity and specificity of the Influ A + B K-SeT® kit were 91.2%-94.5% and 98.2%, respectively, compared to the Xpert® Flu assay and the Simplexa™ Flu A/B & RSV Direct Kit, respectively.

Conclusions: Considering the satisfactory diagnostic performance of the Influ A + B K-SeT® test, its implementation as a point-of-care is expected to aid in decision-making, minimize time-to-result and reduce healthcare costs.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diagmicrobio.2020.115097DOI Listing

Publication Analysis

Top Keywords

influ a + b
20
a + b k-set®
20
performance influ
12
implementation point-of-care
8
k-set® kit
8
simplexa™ flu
8
flu a/b
8
a/b rsv
8
rsv direct
8
direct kit
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!