Introduction: Plasma D-lactate levels can be considered a surrogate indicator of MG flux. There are commercial methods specifically designed to cover pathological ranges (mmol/l in sepsis or leaky gut) that are not suitable to encompass the metabolically relevant ranges.

Material And Methods: We modified and optimized the D-lactate Colorimetric Assay kit MAK058 from Sigma adding an ultracentrifugation step, kinetic reading and increasing reaction temperature.

Results And Conclusions: We present a modified, optimized and validated method to measure serum D-lactate using a commercial kit to increase the sensitivity of the method to 2 orders of magnitude lower as well as minimizing interferences so that it may be used by researchers to explore the D-lactate pathway in metabolic disorders at a low cost (colorimetric method and a plate reader) and with a much higher specificity.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ymeth.2020.06.010DOI Listing

Publication Analysis

Top Keywords

method measure
8
metabolically relevant
8
modified optimized
8
d-lactate
5
optimized sensitive
4
sensitive inexpensive
4
method
4
inexpensive method
4
measure d-lactate
4
d-lactate surrogate
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!