Purpose: To apply failure mode and effect analysis (FMEA) to generate an effective and efficient initial physics plan checklist.

Methods: A team of physicists, dosimetrists, and therapists was setup to reconstruct the workflow processes involved in the generation of a treatment plan beginning from simulation. The team then identified possible failure modes in each of the processes. For each failure mode, the severity (S), frequency of occurrence (O), and the probability of detection (D) was assigned a value and the risk priority number (RPN) was calculated. The values assigned were based on TG 100. Prior to assigning a value, the team discussed the values in the scoring system to minimize randomness in scoring. A local database of errors was used to help guide the scoring of frequency.

Results: Twenty-seven process steps and 50 possible failure modes were identified starting from simulation to the final approved plan ready for treatment at the machine. Any failure mode that scored an average RPN value of 20 or greater was deemed "eligible" to be placed on the second checklist. In addition, any failure mode with a severity score value of 4 or greater was also considered for inclusion in the checklist. As a by-product of this procedure, safety improvement methods such as automation and standardization of certain processes (e.g., dose constraint checking, check tools), removal of manual transcription of treatment-related information as well as staff education were implemented, although this was not the team's original objective. Prior to the implementation of the new FMEA-based checklist, an in-service for all the second checkers was organized to ensure further standardization of the process.

Conclusion: The FMEA proved to be a valuable tool for identifying vulnerabilities in our workflow and processes in generating a treatment plan and subsequently a new, more effective initial plan checklist was created.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7484852PMC
http://dx.doi.org/10.1002/acm2.12918DOI Listing

Publication Analysis

Top Keywords

failure mode
20
analysis fmea
8
fmea generate
8
initial plan
8
workflow processes
8
treatment plan
8
failure modes
8
mode severity
8
failure
7
plan
6

Similar Publications

Purpose: The major cardiac voltage-gated sodium channel Na1.5 (I) is essential for cardiac action potential initiation and subsequent propagation. Compound Chinese medicine Wenxin Keli (WXKL) has been shown to suppress arrhythmias and heart failure.

View Article and Find Full Text PDF

Pretheranostic agents with extraordinaryNIRF/photoacoustic imaging performanceand photothermal oncotherapy efficacy.

Acta Pharm Sin B

December 2024

Guangdong Provincial Key Laboratory of New Drug Screening, NMPA Key Laboratory for Research and Evaluation of Drug Metabolism and Guangdong-Hong Kong-Macao Joint Laboratory for New Drug Screening, School of Pharmaceutical Sciences, Southern Medical University, Guangzhou 510515, China.

Cervical cancer, the most common gynecological malignancy, significantly and adversely affects women's physical health and well-being. Traditional surgical interventions and chemotherapy, while potentially effective, often entail serious side effects that have led to an urgent need for novel therapeutic methods. Photothermal therapy (PTT) has emerged as a promising approach due to its ability to minimize damage to healthy tissue.

View Article and Find Full Text PDF

In order to systematically study the bending behavior of the connection joints of the disc-buckle type formwork support, the accurate numerical model of the disc-buckle type connection joints was obtained through the experimental on the bending behavior of the connection joints of the disc-buckle type, and the bending moment-rotation curve of the joints was verified. The analysis of the failure mode and stress distribution of the joints reveals the importance of the bending behavior of each component. By establishing an accurate numerical model of the joint, the accuracy of the bending experiment of the joint was verified, and the parametric analysis of the influence factors such as the depth of the wedge insertion the disk-plate, the initial position of the wedge insertion the disk-plate, the thickness of the wedge, material constitutive of the wedge and the thickness of the disk-plate was carried out to grasp the influence of the relevant parameters on the bending behavior of the joint.

View Article and Find Full Text PDF

Genetic Assessment of Living Kidney Transplant Donors: A Survey of Canadian Practices.

Can J Kidney Health Dis

January 2025

Multiorgan Transplant Program, Division of Nephrology, Department of Medicine, McGill University Health Centre, Montreal, QC, Canada.

Background: Kidney failure is a prevalent condition with tendency for familial clustering in up to 27% of the affected individuals. Living kidney donor (LKD) transplantation is the optimal treatment option; however, in Canada, more than 45% of LKDs are biologically related to their recipients which subjects recipients to worse graft survival and donors to higher future risk of kidney failure. Although not fully understood, this observation could be partially explained by genetic predisposition to kidney diseases.

View Article and Find Full Text PDF

Background: Simulation offers an opportunity to practice neonatal resuscitation and test clinical systems to improve safety. The authors used simulation-based clinical systems testing (SbCST) with a Healthcare Failure Mode and Effect Analysis (HFMEA) rubric to categorize and quantify latent safety threats (LSTs) during in situ training in eight rural delivery hospitals. The research team hypothesized that most LSTs would be common across hospitals.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!