Introduction: Health care workers are vulnerable to workplace violence, including active shooter incidents. Little is known about how firearms could damage monoplace chamber acrylic and whether a breached pressurized chamber presents additional threat to the patient or bystanders.
Methods: In a remote area where firearm discharge is permitted, we tested the durability of sections of monoplace hyperbaric chamber acrylic under various firearm discharges. Firearms were discharged at acrylic sections from a distance of 17 feet at 45 degrees and 10 degrees from perpendicular while wearing protective gear. Firearm calibers ranged from .22 caliber handgun to 5.56 mm AR-15 rifle. We also conducted similar testing on a monoplace hyperbaric chamber pressurized with >99% oxygen to a differential pressure of 14.7 psig (2.0 atmospheres absolute at sea level). Handguns were remotely fired at a distance of 12 feet from the chamber (30 degrees from perpendicular), while the rifles were fired at a distance of 60 feet from the chamber.
Result: Higher-caliber handguns penetrated or fractured the acrylic sections only after multiple shots. The tested rifles caused full-thickness penetration and fracture with a single shot. However, the pressurized monoplace hyperbaric chamber required two shots from the AR-15 rifle, separated by approximately 60 mm, to penetrate the acrylic, resulting in rapid depressurization. The chamber otherwise remained intact, with no explosion or conflagration observed.
Conclusion: An intact or pressurized chamber performs differently than stand-alone acrylic sections under firearms testing. In a worst-case active shooter scenario, the pressurized monoplace chamber tested posed no additional threat to bystanders beyond the significant risk of ricochet.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.22462/04.06.2020.8 | DOI Listing |
Undersea Hyperb Med
January 2025
Graduate Program in Veterinary Medicine, Center of Rural Science, Federal University of Santa Maria (UFSM), Av. Roraima, 1000, building 42, room 3135, 97105-900 Santa Maria, RS, Brazil.
The assessment of rectal temperature and behavior is an important parameter in all patients for whom hyperbaric oxygen (HBO) therapy is used. The study aims to verify if there is less reduction in body temperature after HBO therapy in restless patients and their behavior during the therapeutic session. Clinical data from 217 HBO therapy sessions with 2 to 2,5 atmospheres absolute (ATA) were reviewed under therapy protocols of 30 (P1) or 45 (P2) minutes, covering 29 canines and 13 felines.
View Article and Find Full Text PDFPLoS One
January 2025
Hyperbaric Medicine Unit, Toronto General Hospital, Toronto, Ontario, Canada.
Background: Hyperbaric oxygen therapy (HBOT) is well established as a treatment for various medical conditions. However, it poses a risk of oxygen toxicity, which can cause seizures particularly in individuals with pre-existing seizure disorders. Consequently, seizure disorders are considered a relative contraindication to HBOT.
View Article and Find Full Text PDFJ Clin Med
November 2024
Department of Emergency Medicine, Yonsei University Wonju College of Medicine, Wonju 26426, Gangwon State, Republic of Korea.
Hyperbaric oxygen therapy (HBOT) involves inhaling nearly 100% oxygen in a pressurized environment and is commonly used to treat various diseases and injuries. Despite its well-known safety, HBOT is associated with side effects, with frequent middle ear barotrauma (MEB) and oxygen toxicity. Understanding the characteristics and risk factors associated with these side effects is critical for improving patient compliance and treatment outcomes.
View Article and Find Full Text PDFAnimals (Basel)
June 2024
Graduate Program in Veterinary Medicine, Center of Rural Science, Federal University of Santa Maria (UFSM), Av. Roraima, 1000, Building 42, Room 3135, Santa Maria 97105-900, RS, Brazil.
Background: This study explored the effects of hyperbaric oxygen therapy (HBOT) on hemogram, serum biochemistry and hemostatic variables in female dogs undergoing laparoscopic-assisted ovariohysterectomy (OVH).
Materials: Thirty adult, mixed-breed, healthy female dogs were randomly divided into the following three groups: HBOT + SURG (exposed to two absolute atmospheres (ATAs) for 45 min followed by laparoscopic-assisted OVH), HBOT (exposed to two ATAs for 45 min) and SURG (laparoscopic-assisted OVH). Blood samples were collected at T0 (at the admission), at T1, 24 h after T0 (immediately after HBOT in the HBOT + SURG and HBOT groups, and immediately before anesthetic premedication in the SURG group), and at T2, 48 h after T0 (24 h after HBOT and anesthetic premedication).
Med Devices (Auckl)
June 2024
Department of Emergency Medicine, Inha University College of Medicine, Incheon, Republic of Korea.
Introduction: A Monoplace hyperbaric chamber delivers oxygen to the patient's tissues through breathing. Gas monitoring inside the chamber is important because oxygen (O) is consumed, and carbon dioxide (CO) is increased because treatment is performed in a closed volume. This study aimed to advance the safety and efficacy of the monoplace hyperbaric chamber (MHC) through mechanical improvement in a gas monitoring system (GMS).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!