Objectives: To assess the effect on impedance levels of intraoperative reinsertion of a cochlear implant (CI) array compared with matched controls.
Study Design: Retrospective patient review.
Setting: Cochlear implant center.
Patients: CI recipients in the Sydney Cochlear Implant Centre (SCIC) database who required intraoperative array reinsertion and matched controls. Exclusion criteria; known preceding meningitis or labyrinthitis ossificans; electrode array buckling; incomplete "final" insertion.
Intervention: Cochlear implantation.
Main Outcome Measures: Impedance values measured intraoperatively, at switch on, 3 months, 6 months, and 12 months postoperatively were analyzed. The Generalized Estimating Equation (GEE) Model was used to compare cases with controls for each device, at each time point, and for each channel.
Results: Thirty-one reinsertion cases identified; six CI 422 arrays; 14 CI 24RE (ST) arrays, and 11 CI 512 arrays. No increase in impedance levels was found in the reinsertion cases when compared with their matched controls. The only statistical difference in impedance was seen in the CI 422 cohort at switch on with the reinsertion cases having lower impedances (p = 0.03).
Conclusion: This is the first study to examine impedance values in patients who underwent intraoperative CI array reinsertion and to compare them with the impedances of matched controls. No significant increase found in impedances between our reinsertion cases and matched controls, suggesting the reinsertion did not result in any additional trauma or inflammation. This has implications for surgery both in routine cases such as a faulty electrode and also for future design of mechanisms for delivery of intracochlear therapies.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MAO.0000000000002650 | DOI Listing |
J Speech Lang Hear Res
January 2025
Aix-Marseille Université, INSERM, INS, Institut de Neurosciences des Systèmes, France.
Purpose: Prelingual deaf children with cochlear implants show lower digit span test scores compared to normal-hearing peers, suggesting a working memory impairment. To pinpoint more precisely the subprocesses responsible for this impairment, we designed a sequence reproduction task with varying length (two to six stimuli), modality (auditory or visual), and compressibility (sequences with more or less regular patterns). Results on 22 school-age children with cochlear implants and 21 normal-hearing children revealed a deficit of children with cochlear implants only in the auditory modality.
View Article and Find Full Text PDFCochlear Implants Int
January 2025
Department of Otorhinolaryngology and Head and Neck Surgery, Erasmus Medical Center, Rotterdam, the Netherlands.
Objective: Early diagnosis, intervention and consistent follow-up of hearing loss is of great importance in children, given the broad impact of untreated childhood hearing loss. Currently, no hearing-specific QOL proxy questionnaire exists for preschool children with hearing loss in the Dutch language. Therefore, the aim of this study was to translate and validate the Preschool HEAR-QL questionnaire into Dutch.
View Article and Find Full Text PDFEar Hear
January 2025
San Francisco Department of Otolaryngology - Head and Neck Surgery, University of California, San Francisco, California, USA.
Objectives: Cochlear implant (CI) user functional outcomes are challenging to predict because of the variability in individual anatomy, neural health, CI device characteristics, and linguistic and listening experience. Machine learning (ML) techniques are uniquely poised for this predictive challenge because they can analyze nonlinear interactions using large amounts of multidimensional data. The objective of this article is to systematically review the literature regarding ML models that predict functional CI outcomes, defined as sound perception and production.
View Article and Find Full Text PDFHear Res
January 2025
CHU Lille, Department of Otology and Neurotology, F-59000 Lille, France; Univ. Lille, France; Univ. Lille, Inserm, CHU Lille, U1008, F-59000 Lille, France.
Objective: The aim of this study is to detail and evaluate the surgical procedure for perilymph sampling from the cochlear apex in the Mongolian gerbil.
Design: Perilymph sampling from the cochlear apex was performed one to three time in 12 male gerbils aged 8 to 12 months via the submandibular route. 11 of them were previously implanted with intracochlear implants loaded with dexamethasone and placed in the scala tympani, the 12th was used to collect control samples.
Int Arch Otorhinolaryngol
January 2025
Department of Otolaryngology, Hospital das Clínicas, Faculty of Medicine, Universidade de São Paulo, São Paulo, SP, Brazil.
Adults with cochlear implants (CIs) need periodic programming of their speech processors to take advantage of alternative adjustments. However, this requires patients to attend the CI center in person. To evaluate the feasibility of speech processor (SP) self-programming with remote assistance in CI users.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!