Purpose: Radiofrequency ablation (RFA) is increasingly being used to treat unresectable liver tumors. Complete ablation of the tumor and a safety margin is necessary to prevent local recurrence. With current electrodes, size and shape of the ablation zone are highly variable leading to unsatisfactory local recurrence rates, especially for tumors >3 cm. In order to improve predictability, we recently developed a system with four simple electrodes with complete ablation in between the electrodes. This rather small but reliable ablation zone is considered as a building block for matrix radiofrequency ablation (MRFA). In the current study we explored the influence of the electric mode (monopolar or bipolar) and the activation mode (consecutive, simultaneous or switching) on the size and geometry of the ablation zone.

Materials And Methods: The four electrode system was applied in ex vivo bovine liver. The electric and the activation mode were changed one by one, using constant power of 50 W in all experiments. Size and geometry of the ablation zone were measured. Finite element method (FEM) modelling of the experiment was performed.

Results: In ex vivo liver, a complete and predictable coagulation zone of a 3 × 2 × 2 cm block was obtained most efficiently in the bipolar simultaneous mode due to the combination of the higher heating efficacy of the bipolar mode and the lower impedance by the simultaneous activation of four electrodes, as supported by the FEM simulation.

Conclusions: In ex vivo liver, the four electrode system used in a bipolar simultaneous mode offers the best perspectives as building block for MRFA. These results should be confirmed by in vivo experiments.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.suronc.2020.02.005DOI Listing

Publication Analysis

Top Keywords

radiofrequency ablation
16
building block
12
vivo liver
12
activation mode
12
size geometry
12
ablation zone
12
ablation
9
ablation electrodes
8
block matrix
8
matrix radiofrequency
8

Similar Publications

"Wound, Pressure Ulcer, and Burn Guidelines-5: Guidelines for the management of lower leg ulcers and varicose veins, second edition" is revised from the first edition, which was published in the Japanese Journal of Dermatology in 2011. The guidelines were drafted by the Wound, Pressure Ulcer, and Burn Guidelines Drafting Committee delegated by the Japanese Dermatological Association and intend to facilitate physicians' clinical decisions in preventing, diagnosing and management of lower leg ulcers and varicose veins. We updated all sections by collecting documents published since the publication of the first edition.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the effectiveness and safety of minimally invasive thermal ablation techniques (microwave ablation and radiofrequency ablation) compared to traditional open surgery for treating benign thyroid nodules.
  • The research included 160 patients treated at a hospital between 2020 and 2023, with a subsequent propensity score matching process to create three comparable groups for analysis.
  • Findings indicate that both microwave and radiofrequency ablation led to less blood loss, shorter surgery times, reduced hospital stays, lower pain levels, fewer complications, and higher patient satisfaction than open surgery.
View Article and Find Full Text PDF

Gastric mixed adenoneuroendocrine carcinoma (MANEC) is a rare and highly aggressive malignancy characterized by both exocrine and neuroendocrine components. Treatment options for metastatic cases are limited, with typical therapeutic approaches involving a combination of chemotherapy and immunotherapy. A 68-year-old male with metastatic gastric MANEC was treated with targeted therapy, immunotherapy, and chemotherapy, including S-1, apatinib, cadonilimab, and paclitaxel.

View Article and Find Full Text PDF

Introduction: Proactive esophageal cooling reduces injury during radiofrequency (RF) ablation of the left atrium (LA) for the treatment of atrial fibrillation (AF). New catheters are capable of higher wattage settings up to 90 W (very high-power short duration, vHPSD) for 4s. Varying power and duration however does not eliminate the risk of thermal injury.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!