AVERAGE GLANDULAR DOSES AND NATIONAL DIAGNOSTIC REFERENCE LEVELS IN MAMMOGRAPHY EXAMINATIONS IN TURKEY.

Radiat Prot Dosimetry

Statistics Department, Art and Science Faculty., Ufuk University, İncek Şehit Savcı Mehmet Selim Kiraz Boulevard, Gölbaşı, Ankara, Turkey.

Published: August 2020

In order to establish national diagnostic reference levels for mammography examinations, entrance skin air kerma, entrance skin dose and average glandular doses (AGDs) were calculated for a total of 25 624 cranio-caudal (CC) and mediolateral oblique (MLO) projections of 6309 patients for 40-49 and 50-64 age groups. The average entrance skin air kerma and entrance skin dose values for both age groups were found to be higher in MLO projections compared with CC projections. The minimum and maximum values of AGDs were determined as 0.4 and 7.9 mGy for both projections. The maximum numbers of AGDs for CC and MLO projections were calculated in the range of 1.1-1.5 and 1.6-2.0 mGy, respectively. The third quartile values of AGDs were calculated for each compressed breast thickness between 20 and 99 mm. The first national diagnostic reference levels of the country were established for each 10-mm compressed breast thickness in mammography examinations.

Download full-text PDF

Source
http://dx.doi.org/10.1093/rpd/ncaa080DOI Listing

Publication Analysis

Top Keywords

entrance skin
16
national diagnostic
12
diagnostic reference
12
reference levels
12
mammography examinations
12
mlo projections
12
average glandular
8
glandular doses
8
levels mammography
8
skin air
8