Purpose: To assess the repeatability and reproducibility of swept-source optical coherence tomography (SS-OCT) and Scheimpflug system and evaluate the agreement between the two systems in measuring multiple corneal regions in children.
Methods: Pachymetric and keratometric maps for both systems were evaluated. Central, midperipheral and peripheral corneal thickness (CT), keratometry and astigmatism power vectors were recorded. The three outcomes yielded by the same observer were used to assess intraobserver repeatability. The differences in the mean values provided by each observer were used to evaluate interobserver reproducibility. Within-subject standard deviation, test-retest repeatability (TRT) and coefficient of variation (CoV) were used to analyze the intraobserver repeatability and interobserver reproducibility. Paired T-test and Bland-Altman were used to appraise interdevice agreement.
Results: Seventy-eight eyes of 78 children were included. The CoV was ≤2.12 and 1.10%, respectively, for repeatability and reproducibility. TRT and CoV were lower for central and paracentral CT measurements than for peripheral measurements. The SS-OCT device generated higher precision when acquiring CT data, whereas Scheimpflug system showed higher reliability when measuring corneal keratometry. Although the CT readings measured using SS-OCT were significantly thinner than Scheimpflug device ( < 0.001), the central and thinnest CT values were still of high agreement. The interdevice agreement of keratometry measurement was high for the central corneal region and moderate for the paracentral and peripheral areas.
Conclusions: The precision of CT measurements by SS-OCT was higher, while the reliability of keratometry measurements by the Scheimpflug system was higher in children. Apart from the measured values in the central corneal region, the thickness and keratometry readings should not be considered interchangeable between the two systems.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7285531 | PMC |
http://dx.doi.org/10.1186/s40662-020-00197-0 | DOI Listing |
BMC Ophthalmol
January 2025
Department of Ophthalmology, The First Affiliated Hospital of Bengbu Medical University, Bengbu, Anhui, 233004, China.
Objectives: To evaluate the effects of short-acting cycloplegic agents, tropicamide and compound tropicamide, on ocular biological parameters and choroid thickness.
Methods: In this study, seventy pediatric subjects aged 6 to 13 years were randomly assigned to two groups: the tropicamide group and compound tropicamide group. Ocular biological parameters and choroidal thickness (CT) and subfoveal choroid thickness (SFCT) were measured in both groups and were retested 40 min after drug administration.
Jpn J Ophthalmol
January 2025
Department of Ophthalmology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita, Osaka, 565-0871, Japan.
Purpose: To determine whether corneal biomechanical parameters can predict ectasia progression.
Study Design: Retrospective observational study.
Methods: The baseline corneal biomechanical parameters of 64 eyes of 41 young patients (age, < 25 years at the first visit) who were diagnosed with keratoconus (KC) or suspected KC at Osaka University Hospital and followed up for more than two years were reviewed.
Front Med (Lausanne)
January 2025
Chongqing Key Laboratory of Ophthalmology, Chongqing Eye Institute, Chongqing Branch (Municipality Division) of National Clinical Research Center for Ocular Diseases, Department of Ophthalmology, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.
Objective: To investigate and evaluate the progression of myopia and associated factors of axial length (AL) growth among children in Chongqing.
Methods: This six-month prospective study was conducted on students in grades 1 to 8 at a school in Chongqing, China. All participants underwent a standard ophthalmologic examination including uncorrected visual acuity (UCVA), noncycloplegic refraction, AL, and corneal topography in March 2023.
Int Med Case Rep J
January 2025
Department of Ophthalmology, University of California San Francisco, San Francisco, CA, USA.
Purpose: To compare the outcomes of two different surgical planning strategies for topography-guided repair of post-LASIK ectasia.
Methods: This is a case report of a patient presenting with post-LASIK ectasia. A retrospective chart review was used to collect details of the ophthalmic exam, as well as ocular imaging such as anterior segment optical coherence tomography and Scheimpflug corneal tomography.
Int J Ophthalmol
January 2025
Department of Medical Surgical Nursing, School of Nursing and Midwifery, Shahid Beheshti University of Medical Sciences, Tehran 1968653111, Iran.
Biomechanical study of the visual system by ocular response analyzer investigates the inter-structural biological relationships, mechanics, and function of the visual system. This review aimed to investigate the changes in corneal biomechanical parameters with age and sex. The articles published in PubMed between 2000 and 2021 were investigated and critiqued, and valid scientific evidence was collected, reviewed and concluded according to the inclusion and exclusion criteria.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!