Our purpose was to evaluate whether the position of a thermoluminescent dosimeter (TLD) crystal results in different exposure readings. Nine subjects wore 2 TLD badges (one facing inward, toward the palm, and one facing outward) for 2 mo. Both TLDs were worn on the middle finger of the dominant hand, with the inward-facing TLD placed at the bottom and the outward-facing TLD at the top. At the end of the first month, these TLDs were replaced with new ones for another month. Combined results from the badges for the 2 mo were recorded in millisieverts. A paired test with 2-sample means was performed to compare the 2 positions in general nuclear medicine and PET/CT subjects, with an α of 0.05. For all subjects and for the general nuclear medicine and PET/CT groups, mean exposure was greater for the inward-facing TLD. For a TLD worn on the dominant hand, extremity-exposure readings are maximized when the TLD faces inward.

Download full-text PDF