AI Article Synopsis

  • Rabies causes about 59,000 deaths globally each year, making rapid diagnosis crucial for quick response and treatment.
  • This study tested 179 brain samples from various rabies-suspected animals using a rapid test kit and a standard fluorescent antibody test, evaluating the accuracy of the rapid test.
  • The rapid kit showed high sensitivity (92%) and perfect specificity (100%), indicating it can effectively confirm rabies cases and support surveillance efforts, especially in areas with limited resources.*

Article Abstract

Background: Rabies kills approximately 59,000 people each year worldwide. Rapid and accurate diagnosis of rabies is important for instituting rapid containment measures and for advising the exposed people for postexposure treatment. The application of a rapid diagnostic tests in the field can greatly enhance disease surveillance and diagnostic activities, especially in resource poor settings. In this study, a total of 179 brain tissue samples collected from different rabies suspect animal species (113 dogs, 50 cattle, 10 cats, 3 goats, 2 horses, and 1 bear) were selected and tested using both rapid immunochromatographic kit and the reference standard fluorescent antibody test (FAT). We evaluated the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of a rapid antigen detection test kit produced by BioNote, Inc. (Hwaseong-si, Korea) relative to a FAT for its fit-for-purpose for confirmation of clinical cases of rabies for early response and enhancing rabies surveillance.

Results: Among 179 samples examined in this study, there was a concordance in results by the rapid test and FAT in 115 positive samples and 54 negative samples. Test results were discordant in 10 samples which were positive by FAT, but negative (false negative) by rapid kit. The rapid test kit showed a sensitivity of 92% (95% CI: 85.9-95.6) and specificity of 100% (95% CI: 93.4-100) using FAT as the reference standard. The positive and negative predictive values were found to be 100% (95% CI:96.7-100) and 84.4% (95% CI: 73.6-91.3), respectively. Overall, there was 94.4% (95% CI: 90-96.9) test agreement between rapid test and FAT (Kappa value = 0.874) with a positive percent agreement and negative percent agreement of 92 and 100%, respectively.

Conclusions: Our finding demonstrated that the rapid test kit (BioNote) can be used for rabies surveillance and confirming clinical case of rabies in animals for making rapid decisions particularly controlling rabies outbreaks in resource poor settings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7281917PMC
http://dx.doi.org/10.1186/s12917-020-02405-4DOI Listing

Publication Analysis

Top Keywords

test kit
16
rapid test
16
test fat
12
rapid
11
test
10
rabies
9
rapid immunochromatographic
8
standard fluorescent
8
fluorescent antibody
8
antibody test
8

Similar Publications

Distinguishing abiotic corrosion from two types of microbiologically influenced corrosion (MIC) using a new electrochemical biofilm/MIC test kit.

J Environ Manage

January 2025

Department of Chemical & Biomolecular Engineering, Institute for Corrosion and Multiphase Technology, Ohio University, Athens, 45701, USA; Department of Biological Sciences, Molecular & Cellular Biology Program, Ohio University, Athens, OH, 45701, USA. Electronic address:

Biofilms can cause biofouling, water quality deterioration, and transmission of infectious diseases. They are also responsible for microbiologically influenced corrosion (MIC) which can cause leaks, resulting in environmental disasters. A new disposable biofilm/MIC test kit was demonstrated to distinguish abiotic corrosion of carbon steel from MIC.

View Article and Find Full Text PDF

Aluminum Induces Neurotoxicity through the MicroRNA-98-5p/Insulin-like Growth Factor 2 Axis.

ACS Chem Neurosci

January 2025

Department of Occupational Health, School of Public Health, Shanxi Medical University, Taiyuan, Shanxi 030001, China.

Aluminum is a well-known and widely distributed environmental neurotoxin. This study aimed to investigate the effect of miR-98-5p targeting insulin-like growth factor 2 (IGF2) on aluminum neurotoxicity. Thirty-two Sprague-Dawley rats were randomly divided into four groups and administered 0, 10, 20, and 40 μmol/kg maltol aluminum [Al(mal)], respectively.

View Article and Find Full Text PDF

Establishing a Three-Dimensional Coculture Module of Epithelial Cells Using Nanofibrous Membranes.

J Vis Exp

December 2024

Department of Pharmacology, School of Medicine, Ajou University; 3D Immune System Imaging Core Center, Ajou University;

Technical hurdles in a culture of epithelial cells include dedifferentiation and loss of function. Biomimetic three-dimensional (3D) cell culture methods can enhance cell culture efficiency. This study introduces an advanced two-layered culture system intended to cultivate epithelial cells as tissue-like layers with the culture of fibroblasts within a 3D environment.

View Article and Find Full Text PDF

Introduction: One of the main causes of primary hepatocellular carcinoma and chronic hepatitis is the hepatitis C virus (HCV), with significant variability in its genotypes affecting pathogenicity and treatment outcomes. In India, prevalence ranges from 0.5 to 1.

View Article and Find Full Text PDF

Background: Diabetes mellitus (DM) is a chronic disease that remains a global health problem with increasing prevalence, particularly in Indonesia. The presence of oral manifestations in patients with DM is influenced by changes in immune system function. Interleukin-10 (IL-10) is an inflammatory marker implicated in DM.

View Article and Find Full Text PDF