  • Fretting corrosion, caused by micromovements in metal joints exposed to fluids, contributes to failures in modular joint arthroplasty. Reducing metal-on-metal contact could help minimize this corrosion risk.
  • The study tested if a ceramic-based silicon nitride (SiNx) coating on taper components affects micromotion in hip prostheses using different metal combinations, applying simulated daily activities to measure micromotion.
  • Findings revealed that Ti6Al4V stems had more micromotion than CoCr29Mo6, and while the SiNx coating did not affect micromotion significantly, it generated wear debris from CoCr29Mo6 heads, indicating challenges in achieving effective separation to reduce in vivo corrosion.

Article Abstract

Fretting corrosion is one contributor to the clinical failure of modular joint arthroplasty. It is initiated by micromotion in metal junctions exposed to fluids. Omitting metal-on-metal contacts could help to reduce the corrosion risk. The coating of one metal taper partner with a ceramic-based silicon nitride (SiNx) coating might provide this separation. The aim of the study was to identify whether a SiNx coating of the male taper component influences the micromotion within a taper junction. Hip prosthesis heads made of CoCr29Mo6 (Aesculap) and Ti6Al4V (Peter Brehm) were assembled (2000 N) to SiNx-coated and uncoated stem tapers made of Ti6Al4V and CoCr29Mo6 (2×2×2 combinations, each n = 4). Consecutive sinusoidal loading representing three daily activities was applied. Contactless relative motion in six degrees of freedom was measured using six eddy-current sensors. Micromotion in the junction was determined by compensating for the elastic deformation derived from additional monoblock measurements. After pull-off, the taper surfaces were microscopically inspected. Micromotion magnitude reached up to 8.4 ± 0.8 µm during loading that represented stumbling. Ti6Al4V stems showed significantly higher micromotion than those made of CoCr29Mo6, while taper coating had no influence. Statistical differences in pull-off forces were found for none of the taper junctions. Microscopy revealed CoCr29Mo6 abrasion from the head taper surface if combined with coated stem tapers. Higher micromotion of Ti6Al4V tapers was probably caused by the lower Young's modulus. Even in the contact areas, the coating was not damaged during loading. The mechanics of coated tapers was similar to uncoated prostheses. Thus, the separation of the two metal surfaces with the objective to reduce in vivo corrosion appears to be achievable if the coating is able to withstand in vivo conditions. However, the hard ceramic-based stem coating lead to undesirable debris from the CoCr29Mo6 heads during loading.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0954411920930616DOI Listing

Publication Analysis

Top Keywords

taper
8
sinx coating
8
stem tapers
8
higher micromotion
8
micromotion
7
coating
7
cocr29mo6
5
sinx coatings
4
coatings bear
4
bear potential
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!