An axial-volume scan with adaptive statistical iterative reconstruction-V (ASIR-V) is newly developed. Our goal was to identify the influence of axial-volume scan and ASIR-V on accuracy of automated nodule volumetry.An "adult' chest phantom containing various nodules was scanned using both helical and axial-volume modes at different dose settings using 256-slice CT. All CT scans were reconstructed using 30% and 50% blending of ASIR-V and filtered back projection. Automated nodule volumetry was performed using commercial software. The image noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) were measured.The axial-volume scan reduced radiation dose by 19.7% compared with helical scan at all radiation dose settings without affecting the accuracy of nodule volumetric measurement (P = .375). Image noise, CNR, and SNR were not significantly different between two scan modes (all, P > .05).The use of axial-volume scan with ASIR-V achieved effective radiation dose reduction while preserving the accuracy of nodule volumetry.