The validation of a method for the determination of total chromium in Fischer-344 rat feces by inductively coupled plasma optical emission spectrometry following a rapid, atmospheric-pressure microwave digestion is described. The performance of the method was evaluated over the concentration range of 5.00 to 200 μg Cr/g feces. Data for method linearity, accuracy, precision, digest stability, and storage stability are presented along with limit of detection and limit of quantitation data. Data from a cross-validation method for B6C3F1 mouse feces are also presented. Following validation, the method was applied to analyze samples collected in support of two chronic toxicological investigations.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7269159 | PMC |
http://dx.doi.org/10.1080/00032710902721931 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!