Background: Finger cuff technologies allow continuous noninvasive arterial blood pressure (AP) and cardiac output/index (CO/CI) monitoring.

Methods: We performed a meta-analysis of studies comparing finger cuff-derived AP and CO/CI measurements with invasive measurements in surgical or critically ill patients. We calculated overall random effects model-derived pooled estimates of the mean of the differences and of the percentage error (PE; CO/CI studies) with 95%-confidence intervals (95%-CI), pooled 95%-limits of agreement (95%-LOA), Cochran's Q and I (for heterogeneity).

Results: The pooled mean of the differences (95%-CI) was 4.2 (2.8 to 5.62) mm Hg with pooled 95%-LOA of -14.0 to 22.5 mm Hg for mean AP (Q=230.4 [P<0.001], I=91%). For mean AP, the mean of the differences between finger cuff technologies and the reference method was ≤5±8 mm Hg in 9/27 data sets (33%). The pooled mean of the differences (95%-CI) was -0.13 (-0.43 to 0.18) L min with pooled 95%-LOA of -2.56 to 2.23 L min for CO (Q=66.7 [P<0.001], I=90%) and 0.07 (0.01 to 0.13) L min m with pooled 95%-LOA of -1.20 to 1.15 L min m for CI (Q=5.8 [P=0.326], I=0%). The overall random effects model-derived pooled estimate of the PE (95%-CI) was 43 (37 to 49)% (Q=48.6 [P<0.001], I=63%). In 4/19 data sets (21%) the PE was ≤30%, and in 10/19 data sets (53%) it was ≤45%.

Conclusions: Study heterogeneity was high. Several studies showed interchangeability between AP and CO/CI measurements using finger cuff technologies and reference methods. However, the pooled results of this meta-analysis indicate that AP and CO/CI measurements using finger cuff technologies and reference methods are not interchangeable in surgical or critically ill patients.

Clinical Trial Number: PROSPERO registration number: CRD42019119266.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.bja.2020.03.013DOI Listing

Publication Analysis

Top Keywords

continuous noninvasive
8
finger cuff
8
cuff technologies
8
arterial blood
8
blood pressure
8
pressure cardiac
8
noninvasive pulse
4
pulse wave
4
wave analysis
4
analysis finger
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!