A demonstration of high field-of-view stability in hands-free echocardiography.

Cardiovasc Ultrasound

Department of Biomedical Engineering, Eindhoven University of Technology, Groene Loper, Building 15, Eindhoven, The Netherlands.

Published: May 2020

Background: Exercise stress echocardiography is clinically used to assess cardiovascular diseases. For accurate cardiac evaluation, a stable field-of-view is required. However, transducer orientation and position are difficult to preserve. Hands-free acquisitions might provide more consistent and reproducible results. In this study, the field-of-view stability and variability of hands-free acquisitions are objectively quantified in a comparison with manually obtained images, based on image structural and feature similarities. In addition, the feasibility and consistency of hands-free strain imaging is assessed.

Methods: In twelve healthy males, apical and parasternal images were acquired hands-free, using a fixation device, and manually, during semi-supine exercise sessions. In the final ten seconds of every exercise period, the image structural similarity and cardiac feature consistency were computed using a steerable pyramid employing complex, oriented wavelets. An algorithm discarding images displaying lung artifacts was created. Hands-free strain consistency was analyzed.

Results: Hands-free acquisitions were possible in 9 of the 12 subjects, whereas manually 10 out of 12 could be imaged. The image structural similarity was significantly improved in the hands-free apical window acquisitions (0.91 versus 0.82), and at least equally good in the parasternal window (0.90 versus 0.82). The change in curvature and orientation of the interventricular septum also appeared to be lower in the hands-free acquisitions. The variability in field-of-view was similar in both acquisitions. Longitudinal, septal strain was shown to be at least as consistent when obtained hands-free compared to manual acquisitions.

Conclusions: The field-of-view was shown to be more or equally stable and consistent in the hands-free data in comparison to manually obtained images. The variability was similar, thus respiration- and exercise-induced motions were comparable for manual and hands-free acquisitions. Additionally, the feasibility of hands-free strain has been demonstrated. Furthermore, the results suggest the hands-free measurements to be more reproducible, though further analysis is required.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7260740PMC
http://dx.doi.org/10.1186/s12947-020-00201-6DOI Listing

Publication Analysis

Top Keywords

hands-free acquisitions
20
hands-free
14
image structural
12
hands-free strain
12
field-of-view stability
8
comparison manually
8
manually images
8
structural similarity
8
versus 082
8
consistent hands-free
8

Similar Publications

Background: In the American Association of Colleges of Nursing's updated Essentials, competency-based education (CBE) is the expected standard for nursing curricula. CBE approaches include observation, practice repetition, demonstration, and video supervision guided by a framework of competency, feedback, and self-reflection.

Method: The promotion of skills competency among nursing students was facilitated by self-regulated learning using video skills-based competency testing.

View Article and Find Full Text PDF

Recent advances in wearable ultrasound technologies have demonstrated the potential for hands-free data acquisition, but technical barriers remain as these probes require wire connections, can lose track of moving targets and create data-interpretation challenges. Here we report a fully integrated autonomous wearable ultrasonic-system-on-patch (USoP). A miniaturized flexible control circuit is designed to interface with an ultrasound transducer array for signal pre-conditioning and wireless data communication.

View Article and Find Full Text PDF

Objective: Advanced driver assistance systems are increasingly available in consumer vehicles, making the study of drivers' behavioral adaptation and the impact of automation beneficial for driving safety. Concerns over driver's being out-of-the-loop, coupled with known limitations of automation, has led research to focus on time-critical, system-initiated disengagements. This study used real-world data to assess drivers' response to, and recovery from, automation-initiated disengagements by quantifying changes in visual attention, vehicle control, and time to steady-state behaviors.

View Article and Find Full Text PDF

Background: Stress echocardiography has been widely used in clinical practice for decades and has recently gained even more importance in diagnostic approaches to ischemic heart disease. However, it still has numerous limitations. Despite advantages of physical exercise as most physiologic stressor, it is impossible to continuously monitor the cardiac function during treadmill test and difficult to maintain an optimal acoustic window during cycle ergometer exercise tests.

View Article and Find Full Text PDF

Purpose: Almost one in four lumpectomies fails to fully remove cancerous tissue from the breast, requiring reoperation. This high failure rate suggests that existing lumpectomy guidance methods are inadequate for allowing surgeons to consistently identify the proper volume of tissue for excision. Current guidance techniques either provide little information about the tumor position or require surgeons to frequently switch between making incisions and manually probing for a marker placed at the lesion site.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!