Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 143
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 143
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 209
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 994
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3134
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
We compared the ability of 2 commercial molecular amplification assays (RealTime SARS-CoV-2 on the 2000 [abbreviated ACOV; Abbott] and ID Now COVID-19 [abbreviated IDNOW; Abbott]) and a laboratory-developed test (modified CDC 2019-nCoV reverse transcriptase PCR [RT-PCR] assay with RNA extraction by eMag [bioMérieux] and amplification on QuantStudio 6 or ABI 7500 real-time PCR system [abbreviated CDC COV]) to detect severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) RNA in upper respiratory tract specimens. Discrepant results were adjudicated by medical record review. A total of 200 nasopharyngeal swab specimens in viral transport medium (VTM) were collected from symptomatic patients between 27 March and 9 April 2020. Results were concordant for 167 specimens (83.5% overall agreement), including 94 positive and 73 negative specimens. The ACOV assay yielded 33 additional positive results, 25 of which were also positive by the CDC COV assay but not by the IDNOW assay. In a follow-up evaluation, 97 patients for whom a dry nasal swab specimen yielded negative results by IDNOW had a paired nasopharyngeal swab specimen collected in VTM and tested by the ACOV assay; SARS-CoV-2 RNA was detected in 13 (13.4%) of these specimens. Medical record review deemed all discrepant results to be true positives. The IDNOW test was the easiest to perform and provided a result in the shortest time but detected fewer cases of COVID-19. The ACOV assay detected more cases of COVID-19 than the CDC COV or IDNOW assays.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7383545 | PMC |
http://dx.doi.org/10.1128/JCM.00938-20 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!