Bevelling of the ultra-fine tips of C1-selective microelectrodes (ME) based on liquid ion-exchanger Corning 477913 caused significant decrease in ME resistivity, increase in sensitivity dependent on its initial value and decrease in selectivity constant for bicarbonates (KC1,HCO3) from 10:1 to approximately 8:1. Bevelling did not affect KC1,NO3 which had value below 1 due to higher selectivity of the ion-exchanger for nitrates. The observed experimental alterations in ME behavior after bevelling were in agreement with those predicted by a relatively simple electrical model comprising of two conductive pathways in parallel having opposite in sign sensitivities.
Download full-text PDF |
Source |
---|
Acta Physiol Pharmacol Bulg
June 1989
Department of Physiology, Medical Academy, Sofia.
Bevelling of the ultra-fine tips of C1-selective microelectrodes (ME) based on liquid ion-exchanger Corning 477913 caused significant decrease in ME resistivity, increase in sensitivity dependent on its initial value and decrease in selectivity constant for bicarbonates (KC1,HCO3) from 10:1 to approximately 8:1. Bevelling did not affect KC1,NO3 which had value below 1 due to higher selectivity of the ion-exchanger for nitrates. The observed experimental alterations in ME behavior after bevelling were in agreement with those predicted by a relatively simple electrical model comprising of two conductive pathways in parallel having opposite in sign sensitivities.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!