Purpose: Between July 2013 and August 2019, 22% of the imaging and radiation oncology core (IROC) spine, and 15% of the moving lung phantom irradiations have failed to meet established acceptability criteria. The spine phantom simulates a highly modulated stereotactic body radiation therapy (SBRT) case, whereas the lung phantom represents a low-to-none modulation moving target case. In this study, we assessed the contribution of dose calculation errors to these phantom results and evaluated their effects on failure rates.
Methods: We evaluated dose calculation errors by comparing the calculation accuracy of various institutions' treatment planning systems (TPSs) vs IROC-Houston's previously established independent dose recalculation system (DRS). Each calculation was compared with the measured dose actually delivered to the phantom; cases in which the recalculation was more accurate were interpreted as a deficiency in the institution's TPS. A total of 258 phantom irradiation plans (172 lung and 86 spine) were recomputed.
Results: Overall, the DRS performed better than the TPSs in 47% of the spine phantom cases. However, the DRS was more accurate in 93% of failing spine phantom cases (with an average improvement of 2.35%), indicating a deficiency in the institution's treatment planning system. Deficiencies in dose calculation accounted for 60% of the overall discrepancy between measured and planned doses among spine phantoms. In contrast, lung phantom DRS calculations were more accurate in only 35% and 42% of all and failing lung phantom cases respectively, indicating that dose calculation errors were not substantially present. These errors accounted for only 30% of the overall discrepancy between measured and planned doses.
Conclusions: Dose calculation errors are common and substantial in IROC spine phantom irradiations, highlighting a major failure mode in this phantom and in clinical treatment management of these cases. In contrast, dose calculation accuracy had only a minimal contribution to failing lung phantom results, indicating that other failure modes drive problems with this phantom and similar clinical treatments.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7686159 | PMC |
http://dx.doi.org/10.1002/mp.14258 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!