Purpose: Idiopathic junctional ectopic tachycardia (JET) is typically refractory to antiarrhythmic agents. Catheter ablation for JET is feasible but is associated with high risk of unintended atrioventricular (AV) block. There is limited data on the appropriate procedural technique and clinical outcomes with catheter ablation for idiopathic JET in adults.
Methods: This is a multicenter, retrospective study of all adult patients (age ≥ 18 years) who underwent catheter ablation for idiopathic JET. Patient, procedural characteristics, and long-term outcomes were evaluated.
Results: Fifteen patients [radiofrequency ablation (RF) = 14 and cryoablation = 1) were treated with catheter ablation. The median age was 58 years with 67% males. All patients underwent mapping of the right atrium and the aortic cusps prior to energy delivery. The location of earliest activation in relation to the atrioventricular (AV) node was postero-superior in 73% (11/15), posterior in 13% (2/15), and superior in 13% (2/15) respectively. Acute success was 100%. Arrhythmia recurrence occurred in 53% (8/15) all of whom underwent a repeat ablation. High-grade AV block requiring permanent pacemaker occurred in 20% (3/15). At 12-month follow-up in the redo-ablation group, 37.5% (3/8) had recurrence of the arrhythmia two of which underwent a third ablation procedure.
Conclusion: Catheter ablation of idiopathic JET in adults is associated with a high rate of recurrence requiring multiple procedures and high risk of AV block requiring a permanent pacemaker. Mapping and ablation of the non-coronary cusp can be considered as the arrhythmia was controlled in 3 patients with no inadvertent AV block.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10840-020-00749-3 | DOI Listing |
Rev Port Cardiol
January 2025
Cardiology Department, Hospital Santa Cruz, Centro Hospitalar Lisboa Ocidental, Lisbon, Portugal.
Introduction And Objectives: Pulmonary vein (PV) isolation is one of the cornerstones of rhythm-control therapy for symptomatic atrial fibrillation (AF) patients. Pulsed field ablation (PFA) is a novel ablation modality that involves the application of electrical pulses causing cellular death, and it has preferential tissue specificity. In this study, we aimed to share a one-year single center experience of AF ablation with PFA.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Department of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
Background: The IntellaNav MiFi OI catheter (MiFi) is equipped with a sensor for local impedance (LI) monitoring and three mini-electrodes. In this study, we investigated the target LI values for a successful pulmonary vein isolation (PVI) under the pacing and ablation technique using the MiFi catheter.
Methods: Twenty-seven patients underwent PVI using the MiFi catheter under mini electrode pacing from the MiFi catheter.
Pacing Clin Electrophysiol
January 2025
Aurora Cardiovascular and Thoracic Services, Aurora Sinai/Aurora St. Luke's Medical Centers, Milwaukee, Wisconsin, USA.
Background: Concurrent Micra leadless pacemaker (Medtronic, Minneapolis, Minnesota) implantation and atrioventricular node (AVN) ablation has been shown to be feasible and safe in patients with symptomatic, drug-refractory atrial fibrillation (AF). However, major complications within the 30 days after concurrent Micra implantation and AVN ablation have been reported. We evaluated the efficacy and safety of the concurrent procedure at our institution.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Department of Cardiology, Yan'an Hospital Affiliated to Kunming Medical University, Kunming, Yunnan, P.R. China.
Objective: The Vizigo sheath, a novel visualizable steerable sheath, has been utilized effectively in the clinical management of atrial fibrillation. However, its application in the ablation of typical atrial flutter (AFL) remains unexplored. This study aims to evaluate and compare the efficacy and safety of the Vizigo sheath against a conventional fixed sheath during catheter ablation for typical AFL.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Section of Laboratory for Animal Experiments, Institute of Medical Science, Medical Research Support Center, Nihon University, School of Medicine, Tokyo, Japan.
Background: Neither the actual in vivo tissue temperatures reached with a novel contact force sensing catheter with a mesh-shaped irrigation tip (TactiFlex SE, Abbott) nor the safety profile has been elucidated.
Methods: In a porcine model (n = 8), thermocouples were implanted epicardially in the superior vena cava, right pulmonary vein, and esophagus close to the inferior vena cava following a right thoracotomy. After chest closure, endocardial ablation was conducted near the thermocouples under fluoroscopic guidance.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!