Objective: Endovascular intervention in uncomplicated type B dissection has not been shown conclusively to confer benefit on patients. The hemodynamic effect of primary entry tear coverage is not known. Endovascular stent grafts were deployed in a model of aortic dissection with multiple fenestrations to study these effects. It is hypothesized that endograft deployment will lead to restoration of parabolic true lumen flow as well as elimination of false lumen flow and transluminal jets and vortices locally while maintaining distal false lumen canalization.

Methods: Thoracic stent grafts were placed in silicone models of aortic dissection with a compliant and mobile intimal flap and installed in a flow loop. Pulsatile fluid flow was established with a custom positive displacement pump, and the models were imaged by four-dimensional flow magnetic resonance imaging. Full flow fields were acquired in the models, and velocities were extracted to calculate flow rates, reverse flow indices, and oscillatory shear index, the last two of which are measures of stagnant and disturbed flows.

Results: Complete obliteration of the false lumen was achieved in grafted aorta, with normal parabolic flow profiles in the true lumen (maximal velocity, 30.4 ± 8.4 cm/s). A blind false lumen pouch was created distal to this with low-velocity (5.8 ± 2.7 cm/s) and highly reversed (27.9% ± 13.9% reverse flow index) flows. In distal free false lumen segments, flows were comparable to ungrafted conditions with maximal velocities on the order of 7.0 ± 2.1 cm/s. Visualization studies revealed forward flow in these regions with left-handed vortices from true to false lumen. Shear calculations in free false lumen regions demonstrated reduced oscillatory shear index.

Conclusions: Per the initial hypothesis, endovascular grafting improved true lumen hemodynamics in the grafted region. Just distally, a prothrombotic flow regimen was noted in the false lumen, yet free false lumen distal to this remained canalized. Clinically, this suggests a need for advancing endovascular intervention beyond sole entry tear coverage to prevent further false lumen canalization through uncovered fenestrations.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2019.06.222DOI Listing

Publication Analysis

Top Keywords

false lumen
44
lumen
14
aortic dissection
12
true lumen
12
flow
12
free false
12
false
11
endovascular intervention
8
entry tear
8
tear coverage
8

Similar Publications

In surgery for acute type A aortic dissection, controlling bleeding from the posterior wall of the proximal anastomosis is particularly challenging. To address this, we use the "reversed turn-up technique." For the reinforcement of the proximal aortic stump, Teflon felt strips were placed inside and outside the suture line with 4-0 polypropylene continuous transverse mattress sutures, and BioGlue was applied to the false lumen.

View Article and Find Full Text PDF

Background: Dialysis Access (DA) stenosis impacts hemodialysis efficiency and patient health, necessitating exams for early lesion detection. Ultrasound is widely used due to its non-invasive, cost-effective nature. Assessing all patients in large hemodialysis facilities strains resources and relies on operator expertise.

View Article and Find Full Text PDF

: Contemporary treatments of acute aortic dissection, including medical, surgical, and endovascular options, are remarkably effective at the management of malperfusion and rupture. Unfortunately, long-term studies indicate that 30%-50% of patients need secondary procedures to treat progressive aneurysmal enlargement of the untreated aorta. The Stent-assisted balloon-induced intimal disruption and relamination in aortic dissection repair (STABILISE) technique was introduced to improve long-term outcomes.

View Article and Find Full Text PDF

A 48-year-old male with a history of hyperlipidemia presented to the emergency department with chest pain. Electrocardiographic abnormalities indicated an acute coronary syndrome. Urgent coronary angiography revealed nondominant right coronary artery (RCA) occlusion.

View Article and Find Full Text PDF

Contrast Agent Pooling in the Descending Aorta Due to Severe Heart Failure.

Clin Pract Cases Emerg Med

November 2024

Mito Kyodo General Hospital, Department of Radiology, University of Tsukuba, Mito, Japan.

Case Presentation: An 86-year-old female presented to our emergency department with chest pain and orthopnea and was diagnosed with heart failure and ST-elevation myocardial infarction, prompting hospitalization. During hospitalization, she developed a fever. A chest and abdominal contrast-enhanced computed tomography (CT), conducted to investigate the cause of the fever, coincidentally revealed sedimentation of contrast agent in the descending aorta.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!