A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Interference of Asfotase Alfa in Immunoassays Employing Alkaline Phosphatase Technology. | LitMetric

Interference of Asfotase Alfa in Immunoassays Employing Alkaline Phosphatase Technology.

J Appl Lab Med

University of East Anglia, Bob Champion Research and Education Building, Norwich, NR4 7UQ, United Kingdom.

Published: March 2020

Background: Asfotase alfa (STRENSIQ®, Alexion Pharmaceuticals, Inc.) is the only approved treatment for patients with pediatric-onset hypophosphatasia, a disease caused by a mutation in the tissue-nonspecific alkaline phosphatase (TNSALP) gene. ALP is often used as signaling system in routine immunoassays. Because asfotase alfa contains the active site of the full ALP enzyme, it can catalyze the substrate as the antibody-conjugated ALP would within an assay. Therefore, its presence in a treated patient's sample may generate false positive or false negative results. We investigated whether the presence of asfotase alfa within a sample induced interference in immunoassays that utilize ALP or alternative detection systems.

Methods: Asfotase alfa was added to samples at concentrations from 0.08-5 µg/mL and analysed on various immunoassays following manufacturer's instructions.

Results: Asfotase alfa was detected in all ALP assays but ALKP1 (RayBiotech). We observed no changes in normetanephrine and noradrenaline (IBL) at any asfotase alfa concentration. However, asfotase alfa notably interfered in an oxytocin (ENZO) assay in nonextracted samples. Extraction using a C18 column eliminated the interference. No interference was observed on automated analyzers using alternative detection system (COBAS fT4 and TSH; Advia Centaur FSH, fT4; Architect LH; FSH). Immulite 2000 fT4, TSH, testosterone and hCG (ALP-based) showed no interference. However, the presence of asfotase alfa resulted in a dose-dependent increase of Troponin I signal.

Conclusion: The presence of asfotase alfa must be taken into consideration when analyzing blood samples in treated patients to avoid any risk of misinterpretation of false positive/negative results. It is essential that assays be tested for this possible interference.

Download full-text PDF

Source
http://dx.doi.org/10.1093/jalm/jfz007DOI Listing

Publication Analysis

Top Keywords

asfotase alfa
40
presence asfotase
12
alfa
10
asfotase
9
alkaline phosphatase
8
alternative detection
8
ft4 tsh
8
interference
6
alp
5
interference asfotase
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!