Introduction: Early diagnosis and appropriate management of neonatal jaundice is crucial in avoiding severe hyperbilirubinemia and brain injury. A low-cost, minimally invasive, point-of-care (PoC) tool for total bilirubin (TB) estimation which can be useful across all ranges of bilirubin values and all settings is the need of the hour.
Objective: To assess the accuracy of Bilistick system, a PoC device, for measurement of TB in comparison with estimation by spectrophotometry.
Design/methods: In this cross-sectional clinical study, in infants who required TB estimation, blood samples in 25-µl sample transfer pipettes were collected at the same time from venous blood obtained for laboratory bilirubin estimation. The accuracy of Bilistick in estimating TB within ±2 mg/dl of bilirubin estimation by spectrophotometry was the primary outcome.
Results: Among the enrolled infants, 198 infants were eligible for study analysis with the mean gestation of 36 ± 2.3 weeks and the mean birth weight of 2368 ± 623 g. The median age at enrollment was 68.5 h (interquartile range: 48-92). Bilistick was accurate only in 54.5% infants in measuring TB within ±2 mg/dl difference of TB measured by spectrophotometry. There was a moderate degree of correlation between the two methods (r = 0.457; 95% CI: 0.339-0.561, p value < 0.001). Bland-Altman analysis showed a mean difference of 0.5 mg/dl (SD ± 4.4) with limits of agreement between -8.2 and +9.1 mg/dl.
Conclusion: Bilistick as a PoC device is not accurate to estimate TB within the clinically acceptable difference (±2 mg/dl) of TB estimation by spectrophotometry and needs further improvement to make it more accurate.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/tropej/fmaa026 | DOI Listing |
JAMA Pediatr
May 2023
Division of Neonatology, Department of Pediatrics, Beatrix Children's Hospital, University Medical Center Groningen, Groningen, the Netherlands.
Neonatology
March 2022
Department of Pediatrics, Beatrix Children's Hospital, University Medical Center Groningen, Groningen, The Netherlands.
Background: Recently, the Bilistick®, a point-of-care instrument to measure bilirubin levels, has been developed. It is fast and cheaper than transcutaneous bilirubin (TCB)-measuring devices, but data on diagnostic properties are scarce.
Objective: This study aimed to compare the performance of the Bilistick® (BM-BS 1.
J Trop Pediatr
December 2020
Department of Neonatology, Fernandez Hospital, Hyderabad, Telangana 500029, India.
Introduction: Early diagnosis and appropriate management of neonatal jaundice is crucial in avoiding severe hyperbilirubinemia and brain injury. A low-cost, minimally invasive, point-of-care (PoC) tool for total bilirubin (TB) estimation which can be useful across all ranges of bilirubin values and all settings is the need of the hour.
Objective: To assess the accuracy of Bilistick system, a PoC device, for measurement of TB in comparison with estimation by spectrophotometry.
EClinicalMedicine
July 2018
Bilimetrix s.r.l., Area Science Park, 34149 Trieste, Italy.
Importance: The real prevalence and clinical burden of severe neonatal jaundice are undefined due to difficulties in measuring total serum bilirubin (TSB) outside secondary and tertiary clinical centers.
Objective: To assess the diagnostic performance of the point-of care Bilistick System (BS) in identifying neonatal jaundice patients requiring treatment.
Design: Between April 2015 and November 2016, 1911 neonates, were recruited to participate in the study.
Pediatr Res
August 2019
Division of Neonatology, Careggi University Hospital of Florence, Florence, Italy.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!